Barricor blood collection tubes are equivalent to PST for a variety of chemistry and immunoassay analytes except for lactate dehydrogenase
- PMID: 31201816
- DOI: 10.1016/j.cca.2019.06.013
Barricor blood collection tubes are equivalent to PST for a variety of chemistry and immunoassay analytes except for lactate dehydrogenase
Abstract
Introduction: The BD Barricor tube uses a novel mechanical separator designed to eliminate gel artifacts, decrease cellular contamination, and improve stability. Here, we evaluated the Barricor tube as a possible replacement for PST using Beckman Coulter analyzers under both optimal, alternative, and suboptimal centrifugation conditions based on BD recommendations.
Methods: Paired PST and Barricor samples were collected from 4 local hospitals and processed based on site-specific preanalytical systems involving automated or manual centrifugation. Centrifugation conditions ranged from 1912 ×g for 10 min (suboptimal), 2060 g for 10 min (alternative), and 4000 ×g for 3 or 10 min (optimal). Tube volume (4.5 vs. 5.5 ml) was also assessed. Forty-three chemistry and immunochemistry analytes were measured on Beckman Coulter DxC and DxI analyzers.
Results: Using an automated preanlaytical system with suboptimal spin conditions, no bias between PST and Barricor was observed for all analytes tested except lactate dehydrogenase (LD). Further investigation revealed significant increase in LD when Barricor was spun for 10 min at 1912, 2060 and 4000 ×g, ranging from +7.4-19.4% vs. PST across the entire measurement interval (87-493 U/l). Smaller tube volume was also associated with higher LD. Differences in LD occurred despite no change in other hemolysis markers such as potassium, phosphate, and AST.
Conclusions: LD is most sensitive to varying centrifugation conditions (time and speed) in Barricor tubes. We recommend that BD centrifugation protocols should be closely evaluated to determine if Barricor is equivalent to PST under local preanalytical configurations.
Keywords: Blood collection; Centrifugation; Lactate dehydrogenase; Preanalytical.
Copyright © 2019 Elsevier B.V. All rights reserved.
Similar articles
-
Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube.Ann Lab Med. 2021 Jan;41(1):114-119. doi: 10.3343/alm.2021.41.1.114. Epub 2020 Aug 25. Ann Lab Med. 2021. PMID: 32829587 Free PMC article.
-
The local clinical validation of a new lithium heparin tube with a barrier: BD Vacutainer® Barricor LH Plasma tube.Biochem Med (Zagreb). 2017 Oct 15;27(3):030706. doi: 10.11613/BM.2017.030706. Epub 2017 Aug 28. Biochem Med (Zagreb). 2017. PMID: 28900369 Free PMC article.
-
The BD Barricor blood collection tube is an acceptable and robust alternative to the PST for use with the Beckman AccuTnI+3 assay.Clin Biochem. 2017 Oct;50(15):851-857. doi: 10.1016/j.clinbiochem.2017.05.015. Epub 2017 Jun 3. Clin Biochem. 2017. PMID: 28583435
-
[Importance of preanalytical step in hemostasis].Ann Biol Clin (Paris). 2013 Jul-Aug;71(4):401-7. doi: 10.1684/abc.2013.0870. Ann Biol Clin (Paris). 2013. PMID: 23906567 Review. French.
-
Blood venous sample collection: Recommendations overview and a checklist to improve quality.Clin Biochem. 2017 Jul;50(10-11):568-573. doi: 10.1016/j.clinbiochem.2017.02.021. Epub 2017 Feb 27. Clin Biochem. 2017. PMID: 28242283 Review.
Cited by
-
A comparison of stability of chemical analytes in plasma from the BD Vacutainer® Barricor™ tube with mechanical separator versus tubes containing gel separator.J Clin Lab Anal. 2020 Feb;34(2):e23060. doi: 10.1002/jcla.23060. Epub 2019 Oct 11. J Clin Lab Anal. 2020. PMID: 31605419 Free PMC article.
-
Comparative study of chemical pathology sample collection tubes at the largest hospital in South Africa.J Med Biochem. 2021 Sep 3;40(4):358-366. doi: 10.5937/jomb0-27216. J Med Biochem. 2021. PMID: 34616225 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Research Materials