Oxygen supplementation in the delivery room: T-piece resuscitator cap open or occluded?
- PMID: 31209277
- DOI: 10.1038/s41372-019-0406-x
Oxygen supplementation in the delivery room: T-piece resuscitator cap open or occluded?
Abstract
Objective: Updated neonatal resuscitation guidelines for free-flow oxygen administration with a T-piece resuscitator recommend higher gas flow and an open T-piece cap. We aimed to determine the effect of gas inflow rate and cap occlusion on oxygen delivery through a T-piece resuscitator.
Study design: Using a NeoPuff™ T-piece, oxygen inflow was randomly adjusted from 4 to 10 liters per minute (LPM). Gas outflow and oxygen concentration were measured with the T-piece cap open and occluded. Data were analyzed with repeated measures 2-way ANOVA.
Result: Gas outflow was significantly decreased with the T-piece cap open compared with occluded at each inflow rate (p < 0.001). There was no difference in oxygen concentration of the outflow gas.
Conclusion: Gas flow from the T-piece is affected by the inflow rate and cap occlusion but oxygen concentration is not. To deliver 5 LPM of free-flow oxygen with the cap open, a minimum of 8 LPM gas inflow is required.
Similar articles
-
Laerdal infant resuscitators are unreliable as free-flow oxygen delivery devices.Am J Perinatol. 1997 Jul;14(6):347-51. doi: 10.1055/s-2007-994158. Am J Perinatol. 1997. PMID: 9217956
-
Can Ambu self-inflating bag and Neopuff infant resuscitator provide adequate and safe manual inflations for infants up to 10 kg weight?Arch Dis Child Fetal Neonatal Ed. 2017 Jul;102(4):F333-F338. doi: 10.1136/archdischild-2016-311830. Epub 2016 Dec 23. Arch Dis Child Fetal Neonatal Ed. 2017. PMID: 28011794
-
A comparison of three neonatal resuscitation devices.Resuscitation. 2005 Oct;67(1):113-8. doi: 10.1016/j.resuscitation.2005.02.016. Resuscitation. 2005. PMID: 16081202
-
Management of Extremely Low Birth Weight Infants in Delivery Room.Clin Perinatol. 2017 Jun;44(2):361-375. doi: 10.1016/j.clp.2017.01.004. Epub 2017 Mar 9. Clin Perinatol. 2017. PMID: 28477666 Review.
-
Oxygen Therapy in the Delivery Room: What Is the Right Dose?Clin Perinatol. 2018 Jun;45(2):293-306. doi: 10.1016/j.clp.2018.01.014. Epub 2018 Mar 7. Clin Perinatol. 2018. PMID: 29747889 Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous