Reagent strips test: A simplified method for prompt analysis of cerebrospinal fluid in neurological disorders in emergency
- PMID: 31211215
- PMCID: PMC6562142
- DOI: 10.1016/j.plabm.2019.e00124
Reagent strips test: A simplified method for prompt analysis of cerebrospinal fluid in neurological disorders in emergency
Erratum in
-
Erratum regarding missing Declaration of Competing Interest statements in previously published articles.Pract Lab Med. 2020 Dec 2;22:e00194. doi: 10.1016/j.plabm.2020.e00194. eCollection 2020 Nov. Pract Lab Med. 2020. PMID: 33376766 Free PMC article.
Abstract
Introduction: The initial examination of cerebrospinal fluid (CSF) for proteins, glucose, and leukocytes, is still the gold standard investigation in some neurological conditions like meningitis. Aims and objective of the present study were to determine the accuracy of urinary reagent strip for semi-quantitative analysis of protein, glucose, leucocytes and erythrocytes in CSF along with its role in emergency for rapid diagnosis of neurological conditions.
Materials and methods: 360 samples of CSF were received in emergency laboratory of a tertiary care hospital in a period of 6 months. All CSF samples were subjected to two types of tests-the definitive test and the index test. CSF microscopy for leucocyte and erythrocyte as well as biochemistry tests for protein and glucose on automated biochemistry analyser were considered as definitive test. The index test for protein, glucose, leucocyte and erythrocyte for the same sample was conducted by Combur-10 urinary reagent strip.
Result: The strip test showed a sensitivity of 99% and a specificity of 54% for proteins. With respect to glucose, the strip was highly sensitive (98%) as well as highly specific (92%).It showed a high sensitivity and specificity for leukocytes ≥ 10 cells/cumm i.e.100% and 96% respectively. For CSF erythrocytes sensitivity and specificity was 100.
Conclusion: Urinary reagent strip can be used routinely for rapid analysis of CSF. If implemented, this technique will be useful especially in emergency settings as well as in areas with limited resources.
Keywords: Accuracy; Cerebrospinal fluid analysis; Emergency setting; Semi-quantitative; Urinary reagent strips.
Figures
References
-
- Adhikary M., Chatterjee R.N. Laboratory evaluation of cases of meningitis attending a tertiary care hospital in India: an observational study. Int J Nutr Pharmacol Neurol Dis. 2013;3:282–288.
-
- Chikkannaiah P., Benachinmardi K.K., Srinivasamurthy V. Semi-quantitative analysis of cerebrospinal fluid chemistry and cellularity using urinary reagent strip: an aid to rapid diagnosis of meningitis. Neurol. India. 2016;64:50–55. - PubMed
-
- Nigrovic L.E., Shah S.S., Neuman M.I. Correction of cerebrospinal fluid protein for the presence of red blood cells in children with a traumatic lumbar puncture. J. Pediatr. 2011;159:158–159. - PubMed
-
- Parmar R.C., Warke S., Sira P., Kamat J.R. Rapid diagnosis of meningitis using reagent strips. Indian J. Med. Sci. 2004;58:62–66. - PubMed
-
- Murthy J.M.K. Tuberculous meningitis: the challenges. Neurol. India. 2010;58:716–722. - PubMed
LinkOut - more resources
Full Text Sources