Expired carbon dioxide: a noninvasive monitor of cardiopulmonary resuscitation
- PMID: 3121209
- DOI: 10.1161/01.cir.77.1.234
Expired carbon dioxide: a noninvasive monitor of cardiopulmonary resuscitation
Abstract
End-tidal CO2 concentration (ETCO2) may serve as a simple noninvasive measurement of the blood flow generated by precordial compression during cardiopulmonary resuscitation (CPR). In a mechanically ventilated porcine preparation of ventricular fibrillation, onset of fibrillation was associated with a rapid decrease in ETCO2 from 4.0 +/- 0.2% to less than 0.7 +/- 0.2%. With precordial compression, it increased to 1.9 +/- 0.3%. Animals that were successfully defibrillated after 12 min of CPR demonstrated an immediate increase in ETCO2. The ETCO2 increased from 1.9 +/- 0.3% to 4.9 +/- 0.3% over an interval of between 30 and 60 sec. These changes in ETCO2 were closely related to proportionally similar decreases and increases in cardiac output (CO), and a close correlation between ETCO2 and CO was demonstrated (r = .92). A similar highly significant correlation between ETCO2 and CO was also demonstrated during open-chest cardiac massage (r = .95). ETCO2 therefore serves as a noninvasive measure of pulmonary blood flow and therefore CO. In 17 successfully resuscitated animals. ETCO2 during precordial compression averaged 1.7 +/- 0.2%, whereas it was only 0.5 +/- 0.1% in five animals in whom resuscitation procedures were unsuccessful (p less than .001). Accordingly, ETCO2 prognosticates outcome during CPR and immediately identifies restoration of spontaneous circulation.