Sensitivity and specificity of viral immunoglobulin M determination by indirect enzyme-linked immunosorbent assay
- PMID: 3125220
- PMCID: PMC266277
- DOI: 10.1128/jcm.26.2.328-332.1988
Sensitivity and specificity of viral immunoglobulin M determination by indirect enzyme-linked immunosorbent assay
Abstract
Four sources of error associated with virus-specific immunoglobulin M (IgM) determination by indirect enzyme-linked immunosorbent assay were recognized and analyzed. First, competitive inhibition due to specific IgG was demonstrated by experiments involving addition and subtraction of rubella-specific IgG. Second, the interference due to rheumatoid factors (RFs) of the IgM class (IgM-RFs) was studied thoroughly, and it appeared that the level of false positivity was more dependent on specific IgG titers than on IgM-RF titers. Third, it was found that some IgM-RFs, differing from conventional IgM-RFs in that they reacted only with isologous IgG, were responsible for further cases of false positivity. Fourth, the interference of an IgM reacting with some virus-unmasked cellular antigens was demonstrated for some uninfected individuals. All four interfering factors could be readily eliminated by simply premixing serum samples with a sheep anti-human gamma-chain serum. This single pretreatment was shown to eliminate false-negatives as well as false-positives in a further 2,004 sera tested for six viruses. These results also emphasize the frequency of RFs and their heterogeneity.