Clinical evaluation of performance, biocompatibility, and safety of vitamin E-bonded polysulfone membrane hemodialyzer compared to non-vitamin E-bonded hemodialyzer
- PMID: 31254226
- PMCID: PMC6861196
- DOI: 10.1007/s10047-019-01110-w
Clinical evaluation of performance, biocompatibility, and safety of vitamin E-bonded polysulfone membrane hemodialyzer compared to non-vitamin E-bonded hemodialyzer
Abstract
The vitamin E-bonded polysulfone membrane hemodialyzer (ViE™-21) was evaluated in a clinical study for regulatory submission. Seventeen patients on hemodialysis were treated with conventional high-flux hemodialyzers for 2 weeks (Pre-ViE phase) and switched to the ViE-21 for 36 sessions (ViE phase) followed by an additional 2 weeks on conventional hemodialyzers (Post-ViE phase). Reduction ratios of urea, creatinine, beta-2-microglobulin, albumin, and ultrafiltration coefficients (KUF) were measured once during the Pre-ViE phase and twice during the ViE phase. Moreover, biocompatibility markers [leucocyte count, platelet count, and activated complement factor (C3a) levels] were evaluated pre-dialysis, 15 min after initiation, and post-dialysis. During the study, type and number of adverse events (AEs), and device malfunctions were recorded. ViE-21 reduction ratios and KUF were not noticeably different than those of conventional hemodialyzers. Fluctuations of leucocyte counts and C3a concentrations were similar using ViE-21 and conventional hemodialyzers; however, the platelet count fluctuation was lower in ViE-21 sessions. The frequency of episodes of hypotension occurring during the ViE phase was lower than that occurring during the Pre- and Post-ViE phases. In conclusion, this study provided performance and safety data of the ViE-21 for regulatory application. The data suggest that vitamin E-bonded hemodialyzers are beneficial in lowering platelet activation and frequency of intradialytic hypotension. Larger randomized controlled trials are needed to confirm these findings.
Keywords: Adverse event; Good clinical practice; Hemodialysis; Intradialytic hypotension; Platelet.
Conflict of interest statement
The study was financially supported by Asahi Kasei Medical Co., Ltd. M. Aritomi and M. Nagase are employees of Asahi Kasei Medical Co., Ltd. There are no other conflicts of interests for this study.
Figures

References
Publication types
MeSH terms
Substances
LinkOut - more resources
Medical
Research Materials