. 2019 Dec;22(4):307-315.
doi: 10.1007/s10047-019-01110-w. Epub 2019 Jun 26.

Clinical evaluation of performance, biocompatibility, and safety of vitamin E-bonded polysulfone membrane hemodialyzer compared to non-vitamin E-bonded hemodialyzer

Affiliations

Clinical evaluation of performance, biocompatibility, and safety of vitamin E-bonded polysulfone membrane hemodialyzer compared to non-vitamin E-bonded hemodialyzer

Mercedeh Kiaii et al. J Artif Organs. 2019 Dec.

Abstract

The vitamin E-bonded polysulfone membrane hemodialyzer (ViE™-21) was evaluated in a clinical study for regulatory submission. Seventeen patients on hemodialysis were treated with conventional high-flux hemodialyzers for 2 weeks (Pre-ViE phase) and switched to the ViE-21 for 36 sessions (ViE phase) followed by an additional 2 weeks on conventional hemodialyzers (Post-ViE phase). Reduction ratios of urea, creatinine, beta-2-microglobulin, albumin, and ultrafiltration coefficients (KUF) were measured once during the Pre-ViE phase and twice during the ViE phase. Moreover, biocompatibility markers [leucocyte count, platelet count, and activated complement factor (C3a) levels] were evaluated pre-dialysis, 15 min after initiation, and post-dialysis. During the study, type and number of adverse events (AEs), and device malfunctions were recorded. ViE-21 reduction ratios and KUF were not noticeably different than those of conventional hemodialyzers. Fluctuations of leucocyte counts and C3a concentrations were similar using ViE-21 and conventional hemodialyzers; however, the platelet count fluctuation was lower in ViE-21 sessions. The frequency of episodes of hypotension occurring during the ViE phase was lower than that occurring during the Pre- and Post-ViE phases. In conclusion, this study provided performance and safety data of the ViE-21 for regulatory application. The data suggest that vitamin E-bonded hemodialyzers are beneficial in lowering platelet activation and frequency of intradialytic hypotension. Larger randomized controlled trials are needed to confirm these findings.

Keywords: Adverse event; Good clinical practice; Hemodialysis; Intradialytic hypotension; Platelet.

PubMed Disclaimer

Conflict of interest statement

The study was financially supported by Asahi Kasei Medical Co., Ltd. M. Aritomi and M. Nagase are employees of Asahi Kasei Medical Co., Ltd. There are no other conflicts of interests for this study.

Figures

Fig. 1
Fig. 1
Fluctuation of biocompatibility markers during the dialysis sessions for the intent-to-treat (ITT) population. Figures a, b, and c correspond to fluctuations of leucocyte, platelet, and C3a, respectively. Open circles, filled circles, and filled squares indicate measurements at the Pre-ViE phase, ViE phase 1st measurement, and ViE phase 2nd measurement, respectively. The 15 min after initiation and post-dialysis values were first normalized by defining the pre-dialysis values as baseline (100%) and using the hematocrit correction equation. The changes (%) at 15 min and post-dialysis are shown in the figure by indicating the pre-dialysis point as 0%. The bars indicate standard errors and P values indicate the results of paired t test compared to the Pre-ViE phase

References

    1. Bowry SK. Nano-controlled membrane spinning technology: regulation of pore size, distribution and morphology of a new polysulfone dialysis membrane. Contrib Nephrol. 2002;137:85–94. doi: 10.1159/000060234. - DOI - PubMed
    1. Krane V, Krieter DH, Olschewski M, et al. Dialyzer membrane characteristics and outcome of patients with type 2 diabetes on maintenance hemodialysis. Am J Kidney Dis. 2007;49:267–275. doi: 10.1053/j.ajkd.2006.11.026. - DOI - PubMed
    1. Del Vecchio L, Locatelli F, Carini M. What we know about oxidative stress in patients with chronic kidney disease on dialysis-clinical effects, potential treatment, and prevention. Semin Dial. 2011;24:56–64. doi: 10.1111/j.1525-139X.2010.00819.x. - DOI - PubMed
    1. Galli F. Vitamin E-modified dialyzers. Contrib Nephrol. 2002;137:95–105. doi: 10.1159/000060235. - DOI - PubMed
    1. Floridi A, Piroddi M, Pilolli F, Matsumoto Y, Aritomi M, Galli F. Analysis method and characterization of the antioxidant capacity of vitamin E-interactive polysulfone hemodialyzers. Acta Biomater. 2009;5:2974–2982. doi: 10.1016/j.actbio.2009.04.011. - DOI - PubMed