Different femoral rotation with navigated flexion-gap balanced or measured resection in total knee arthroplasty does not lead to different clinical outcomes

Affiliations

Different femoral rotation with navigated flexion-gap balanced or measured resection in total knee arthroplasty does not lead to different clinical outcomes

José A Hernández-Hermoso et al. Knee Surg Sports Traumatol Arthrosc. 2020 Jun.

Abstract

Purpose: Femoral rotation in total knee arthroplasty (TKA) is hypothesized to vary in the same knee depending on the method used to establish it.

Methods: Thirty-eight patients who underwent TKA surgery using a measured resection technique (RT) were compared with 40 patients who underwent a flexion-gap balancing technique with computer-assisted (for navigation) surgery (FB-CAS) to assess clinical and radiographic alignment differences at two years postoperatively. In 36 of the 40 patients in the FB-CAS group, both methods were used. Intraoperatively, the transepicondylar femoral rotation (TEFR) in reference to the transepicondylar axis was established as the rotation that balanced the flexion gap. Once the TEFR was obtained, an analogous rotation as measured by a posterior reference femoral rotation (PRFR) cutting guide was determined.

Results: Femoral component rotation determined by the TEFR and PRFR methods differed in each of the knees. The median TEFR was 0.08°±0.6° (range - 1.5°, 1.5°), and the median PRFR was 0.06°±2.8° (range - 6°, 5°). The mean difference in the rotational alignment between the TEFR and PRFR techniques was 0.01° ± 3.1°. The 95% limits of agreement between the mean differences in measurements were between 6.2° external rotation and - 6.1° internal rotation. At 2 years postoperatively, we found no differences in the radiographic or clinical American Knee Society score between the two groups.

Conclusion: Rotation of the femoral component in TKA can vary in the same knee depending on the surgical method used to establish it. This variation in femoral rotation is sufficiently small enough to have no apparent effect on the 2-year clinical outcome score.

Level of evidence: II.

Keywords: Femoral rotation; Gap balancing; Measured resection; Navigation; Total knee arthroplasty; Transepicondylar axis.

PubMed Disclaimer

References

    1. Berger RA, Rubasch HE, Seel MJ, Warren HT, Crosset LS (1993) Determining the rotational alignment of the femoral component in total knee arthroplasty using the epicondylar axis. Clin Orthop 286:40–47
    1. Bland JM, Altman DG (1999) Measuring agreement in method comparison studies. Stat Methods Med Res 8:135–160 - DOI - PubMed
    1. Chen Z, Wang L, Liu Y, He J, Lian Q, Li D, Jin Z (2015) Effect of component mal-rotation on the knee loading in total knee arthroplasty using multi-body dynamics modeling under a simulated walking gait. J Orthop Res 33(9):1287–1296 - DOI - PubMed
    1. Cheng T, Zhao S, Peng X, Zhang X (2012) Does computer assisted surgery improve postoperative leg alignment and implant positioning following total knee arthroplasty? A meta-analysis of randomized control trials. Knee Surg Sports Traumatol Arthrosc 20:1307–1322 - DOI - PubMed
    1. Chon JG, Sun DH, Jung JY, Kim TI, Jnag SW (2011) Rotational alignment of the femoral component for minimal medial collateral ligament release in total knee arthroplasty. Knee Surg Relat Res 23(3):153–158 - DOI - PubMed - PMC

LinkOut - more resources