Accuracy of an indirect carbon dioxide Fick method in determination of the cardiac output in critically ill mechanically ventilated patients
- PMID: 3129478
- DOI: 10.1007/BF00257466
Accuracy of an indirect carbon dioxide Fick method in determination of the cardiac output in critically ill mechanically ventilated patients
Abstract
We evaluated the accuracy of an indirect CO2 Fick method for measuring cardiac output in 30 critically ill mechanically ventilated patients. When the Fick principle was applied to CO2 using estimated PaCO2, the cardiac output obtained underestimated the thermodilution technique showing a lack of accuracy. However, there was a significant correlation between thermodilution and CO2 rebreathing methods using measured (r = 0.92; p less than 0.001) and estimated (r = 0.60; p less than 0.01) arterial PCO2. The regression equation using measured arterial PCO2 was y = 0.59 + 0.91x, and for estimated arterial PCO2 was y = 1.7 + 0.33x. The results suggest that the CO2 rebreathing method using measured arterial PCO2 may be useful to determine cardiac output in those seriously ill patients on artificial ventilation not requiring right heart catheterization.
Similar articles
-
Carbon dioxide rebreathing method of cardiac output measurement during acute respiratory failure in patients with chronic obstructive pulmonary disease.Crit Care Med. 1994 Jan;22(1):81-5. doi: 10.1097/00003246-199401000-00017. Crit Care Med. 1994. PMID: 8124980
-
Non-invasive measurement of cardiac output in patients with acute lung injury using the carbon dioxide rebreathing method.Clin Intensive Care. 1994;5(4):172-5. Clin Intensive Care. 1994. PMID: 10150542
-
Oxygen Fick and modified carbon dioxide Fick cardiac outputs.Crit Care Med. 1994 Jan;22(1):86-95. doi: 10.1097/00003246-199401000-00018. Crit Care Med. 1994. PMID: 8124981
-
Estimating Arterial Partial Pressure of Carbon Dioxide in Ventilated Patients: How Valid Are Surrogate Measures?Ann Am Thorac Soc. 2017 Jun;14(6):1005-1014. doi: 10.1513/AnnalsATS.201701-034FR. Ann Am Thorac Soc. 2017. PMID: 28570147 Review.
-
Measurement of cardiac output by carbon dioxide rebreathing methods.Clin Chest Med. 1989 Jun;10(2):255-64. Clin Chest Med. 1989. PMID: 2500297 Review.
Cited by
-
Noninvasive cardiac output determined with a new method based on gas exchange measurements and carbon dioxide rebreathing: a study in animals/pigs.J Clin Monit. 1992 Oct;8(4):267-78. doi: 10.1007/BF01617908. J Clin Monit. 1992. PMID: 1453186
-
Use of capnography to detect hypercapnic episodes during weaning from mechanical ventilation.Intensive Care Med. 1996 May;22(5):374-81. doi: 10.1007/BF01712151. Intensive Care Med. 1996. PMID: 8796386 Clinical Trial.
-
Non-invasive pulmonary blood flow measurement by means of CO2 analysis of expiratory gases.Intensive Care Med. 1991;17(2):98-102. doi: 10.1007/BF01691431. Intensive Care Med. 1991. PMID: 1907622
-
Comparison of a modified Fick method with thermodilution for determining cardiac output in critically ill patients on mechanical ventilation.Intensive Care Med. 1990;16(4):248-51. doi: 10.1007/BF01705160. Intensive Care Med. 1990. PMID: 2358558
-
Carbon dioxide kinetics and capnography during critical care.Crit Care. 2000;4(4):207-15. doi: 10.1186/cc696. Epub 2000 Jul 12. Crit Care. 2000. PMID: 11094503 Free PMC article. Review.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Other Literature Sources