Accuracy of a portable pulse oximeter in monitoring hypoxemic infants with cyanotic heart disease
- PMID: 31304897
- DOI: 10.1017/S1047951119001355
Accuracy of a portable pulse oximeter in monitoring hypoxemic infants with cyanotic heart disease
Abstract
Objective: Infants with single ventricle physiology have arterial oxygen saturations between 75 and 85%. Home monitoring with daily pulse oximetry is associated with improved interstage survival. They are typically sent home with expensive, bulky, hospital-grade pulse oximeters. This study evaluates the accuracy of both the currently used Masimo LNCS and a relatively inexpensive, portable, and equipped with Bluetooth technology study device, by comparing with the gold standard co-oximeter.
Design: Prospective, observational study.
Setting: Single institution, paediatric cardiac critical care unit, and neonatal ICU.
Interventions: none.
Patients: Twenty-four infants under 12 months of age with baseline oxygen saturation less than 90% due to cyanotic CHD.
Measurements and results: Pulse oximetry with WristOx2 3150 with infant sensors 8008 J (study device) and Masimo LCNS saturation sensor connected to a Philips monitor (hospital device) were measured simultaneously and compared to arterial oxy-haemoglobin saturation measured by co-oximetry. Statistical analysis evaluated the performances of each and compared to co-oximetry with Schuirmann's TOST equivalence tests, with equivalence defined as an absolute difference of 5% saturation or less. Neither the study nor the hospital device met the predefined standard for equivalence when compared with co-oximetry. The study device reading was on average 4.0% higher than the co-oximeter, failing to show statistical equivalence (p = 0.16). The hospital device was 7.4% higher than the co-oximeter and also did not meet the predefined standard for equivalence (p = 0.97).
Conclusion: Both devices tended to overestimate oxygen saturation in this patient population when compared to the gold standard, co-oximetry. The study device is at least as accurate as the hospital device and offers the advantage of being more portable with Bluetooth technology that allows reliable, efficient data transmission. Currently FDA-approved, smaller portable pulse oximeters can be considered for use in home monitoring programmes.
Keywords: Pulse oximetry; congenital heart disease/defects; home monitoring; hypoxia; monitoring; paediatric.
Similar articles
-
Accuracy of the Masimo SET® LNCS neo peripheral pulse oximeter in cyanotic congenital heart disease.Cardiol Young. 2016 Aug;26(6):1183-6. doi: 10.1017/S1047951115002188. Epub 2015 Oct 16. Cardiol Young. 2016. PMID: 26471067
-
Accuracy of Pulse Oximeters Intended for Hypoxemic Pediatric Patients.Pediatr Crit Care Med. 2016 Apr;17(4):315-20. doi: 10.1097/PCC.0000000000000660. Pediatr Crit Care Med. 2016. PMID: 26914626
-
Clinical use of new-generation pulse oximeters in the neonatal intensive care unit.Am J Perinatol. 2005 Oct;22(7):357-60. doi: 10.1055/s-2005-872704. Am J Perinatol. 2005. PMID: 16215921
-
From oxygen content to pulse oximetry: completing the picture in the newborn.Acta Anaesthesiol Scand Suppl. 1995;107:95-100. doi: 10.1111/j.1399-6576.1995.tb04341.x. Acta Anaesthesiol Scand Suppl. 1995. PMID: 8599308 Review.
-
Racial and skin color mediated disparities in pulse oximetry in infants and young children.Paediatr Respir Rev. 2024 Jun;50:62-72. doi: 10.1016/j.prrv.2023.12.006. Epub 2024 Jan 5. Paediatr Respir Rev. 2024. PMID: 38233229 Free PMC article. Review.
Cited by
-
Melanometry for objective evaluation of skin pigmentation in pulse oximetry studies.Commun Med (Lond). 2024 Jul 11;4(1):138. doi: 10.1038/s43856-024-00550-7. Commun Med (Lond). 2024. PMID: 38992188 Free PMC article. Review.
-
Pulse Oximetry Error in Patients With Single-Ventricle Palliation.J Soc Cardiovasc Angiogr Interv. 2022 Jun 30;1(5):100394. doi: 10.1016/j.jscai.2022.100394. eCollection 2022 Sep-Oct. J Soc Cardiovasc Angiogr Interv. 2022. PMID: 39131470 Free PMC article. No abstract available.
-
Wireless monitoring devices in hospitalized children: a scoping review.Eur J Pediatr. 2023 May;182(5):1991-2003. doi: 10.1007/s00431-023-04881-w. Epub 2023 Mar 2. Eur J Pediatr. 2023. PMID: 36859727 Free PMC article.
-
Effect of skin tone on the accuracy of the estimation of arterial oxygen saturation by pulse oximetry: a systematic review.Br J Anaesth. 2024 May;132(5):945-956. doi: 10.1016/j.bja.2024.01.023. Epub 2024 Feb 17. Br J Anaesth. 2024. PMID: 38368234 Free PMC article.
-
The Accuracy of Readily Available Consumer-Grade Oxygen Saturation Monitors in Pediatric Patients.Respir Care. 2024 Mar 27;69(4):387-394. doi: 10.4187/respcare.11258. Respir Care. 2024. PMID: 38164568 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous