. 2019 Dec;139(3):185-195.
doi: 10.1007/s10633-019-09707-5. Epub 2019 Jul 16.

Test-retest repeatability of the pattern electroretinogram and flicker electroretinogram

Affiliations

Test-retest repeatability of the pattern electroretinogram and flicker electroretinogram

Arthur F Resende et al. Doc Ophthalmol. 2019 Dec.

Abstract

Purpose: To evaluate the repeatability of the steady-state pattern electroretinogram (PERG) and full-field flicker electroretinogram (Flicker ERG) protocols, delivered by the office-based Neuro Optic Vision Assessment (NOVA)™ testing platform, in healthy subjects.

Methods: Healthy individuals underwent PERG (16° and 24°) and Flicker ERG [fixed luminance (FL) and multi-luminance (ML)] testing protocols. Test-retest repeatability of protocols was calculated using intra-class correlation coefficients (ICC). Reference values of the parameters of the aforementioned tests were also calculated.

Results: The ICCs for the PERG parameters ranged from 0.793 to 0.911 (p < 0.001). The ICCs for the Flicker ERG parameters ranged from 0.968 to 0.994 (p < 0.001). A linear regression analysis was applied to assess the impact of age on ERG responses. Age had a significant impact on all PERG parameters (16° or 24°). The phase response of the FL Flicker ERG significantly decreased with age (β = - 0.837, p ≤ 0.001). The FL Flicker ERG Magnitude was also impacted with a significant quadratic effect of age (β = - 0.0047, p = 0.0004). Similarly, the Phase Area Under the Curve (Phase AUC) of the ML Flicker ERG significantly declined with age (β = - 0.007, p = 0.009), and the impact on the Magnitude AUC was significant as well, with a negative quadratic age effect.

Conclusions: The PERG and Flicker ERG protocols, delivered by an office-based testing platform, were shown to have good-to-excellent test-retest repeatability when tests were performed in the same order and in immediate succession.

Keywords: Electrophysiology; Flicker electroretinogram; Reference database; Repeatability; Steady-state pattern electroretinogram.

PubMed Disclaimer

References

    1. Doc Ophthalmol. 2014 Oct;129(2):129-38 - PubMed
    1. Ophthalmology. 2008 Jun;115(6):957-63 - PubMed
    1. J Diabetes Res. 2015;2015:319692 - PubMed
    1. J Glaucoma. 2009 Aug;18(6):437-42 - PubMed
    1. Invest Ophthalmol Vis Sci. 2013 Mar 28;54(3):2346-52 - PubMed

Publication types

LinkOut - more resources