Comparative Study
. 2019 Nov:126:103907.
doi: 10.1016/j.mvr.2019.103907. Epub 2019 Jul 19.

Quantitative assessment of the conjunctival microcirculation using a smartphone and slit-lamp biomicroscope

Affiliations
Comparative Study

Quantitative assessment of the conjunctival microcirculation using a smartphone and slit-lamp biomicroscope

Paul F Brennan et al. Microvasc Res. 2019 Nov.

Abstract

Purpose: The conjunctival microcirculation is a readily-accessible vascular bed for quantitative haemodynamic assessment and has been studied previously using a digital charge-coupled device (CCD). Smartphone video imaging of the conjunctiva, and haemodynamic parameter quantification, represents a novel approach. We report the feasibility of smartphone video acquisition and subsequent haemodynamic measure quantification via semi-automated means.

Methods: Using an Apple iPhone 6 s and a Topcon SL-D4 slit-lamp biomicroscope, we obtained videos of the conjunctival microcirculation in 4 fields of view per patient, for 17 low cardiovascular risk patients. After image registration and processing, we quantified the diameter, mean axial velocity, mean blood volume flow, and wall shear rate for each vessel studied. Vessels were grouped into quartiles based on their diameter i.e. group 1 (<11 μm), 2 (11-16 μm), 3 (16-22 μm) and 4 (>22 μm).

Results: From the 17 healthy controls (mean QRISK3 6.6%), we obtained quantifiable haemodynamics from 626 vessel segments. The mean diameter of microvessels, across all sites, was 21.1μm (range 5.8-58 μm). Mean axial velocity was 0.50mm/s (range 0.11-1mm/s) and there was a modestly positive correlation (r 0.322) seen with increasing diameter, best appreciated when comparing group 4 to the remaining groups (p < .0001). Blood volume flow (mean 145.61pl/s, range 7.05-1178.81pl/s) was strongly correlated with increasing diameter (r 0.943, p < .0001) and wall shear rate (mean 157.31 s-1, range 37.37-841.66 s-1) negatively correlated with increasing diameter (r - 0.703, p < .0001).

Conclusions: We, for the first time, report the successful assessment and quantification of the conjunctival microcirculatory haemodynamics using a smartphone-based system.

Keywords: Conjunctival circulation; Endothelial dysfunction; Haemodynamic assessment; Microcirculation; Smartphone.