Comparative Study
. 2020 Jan;39(1):77-83.
doi: 10.1097/ICO.0000000000002052.

Comparison of Corneal Power and Corneal Astigmatism of Different Diameter Zones Centered on the Pupil and Corneal Apex Using Scheimpflug Tomography

Affiliations
Comparative Study

Comparison of Corneal Power and Corneal Astigmatism of Different Diameter Zones Centered on the Pupil and Corneal Apex Using Scheimpflug Tomography

Jing Dong et al. Cornea. 2020 Jan.

Abstract

Purpose: Using Scheimpflug tomography to investigate the difference in corneal power and corneal astigmatism between 3- and 4-mm diameter zones centered on the pupil and corneal apex.

Methods: A total of 90 eyes were included in this study. Axial keratometry, total refractive power, and true net power centered on the pupil and corneal apex in 3- and 4-mm diameter zones were assessed. The paired sample t test and independent sample t test were used for data comparison.

Results: For corneal power and corneal astigmatism for the 3- and 4-mm diameter zones, the flat K values of axial keratometry, total refractive power, and true net power centered on the pupil (3-mm: 43.99 ± 1.69 D, 43.12 ± 1.71 D, 42.53 ± 1.67 D; 4-mm: 44.04 ± 1.67 D, 43.38 ± 1.71 D, 42.61 ± 1.65 D) were significantly higher than those centered on the apex (3-mm: 43.93 ± 1.69 D, 43.05 ± 1.71 D, 42.46 ± 1.67 D; 4-mm: 44.01 ± 1.67 D, 43.34 ± 1.69 D, 42.58 ± 1.65 D; all P < 0.02). However, the steep K and astigmatism magnitude values centered on the pupil (3-mm: 45.71 ± 1.73 D, 45.01 ± 1.73 D, 44.38 ± 1.69 D; 1.72 ± 0.83 D, 1.89 ± 0.86 D, 1.85 ± 0.84 D; 4-mm: 45.78 ± 1.73 D, 45.28 ± 1.74 D, 44.45 ± 1.68 D; 1.73 ± 0.84 D, 1.90 ± 0.85 D, 1.84 ± 0.85 D) were lower than those centered on the apex (3-mm: 45.81 ± 1.74 D, 45.10 ± 1.72 D, 44.50 ± 1.70 D; 1.88 ± 0.90 D, 2.05 ± 0.90 D, 2.04 ± 0.90 D; 4-mm: 45.85 ± 1.73 D, 45.34 ± 1.73 D, 44.51 ± 1.69 D; 1.83 ± 0.88 D, 2.00 ± 0.90 D, 1.93 ± 0.86 D; all P < 0.01). Compared with the 3-mm diameter zone, the corresponding 4-mm diameter zone showed higher corneal power centered on both pupil and corneal apex.

Conclusions: The difference in corneal power and corneal astigmatism, measured at 3 and 4 mm, centered on the pupil and the corneal apex should be noted in corneal refractive surgery and toric intraocular lens calculation.

PubMed Disclaimer

References

    1. Wachler BS, Korn TS, Chandra NS, et al. Decentration of the optical zone: centering on the pupil versus the coaxially sighted corneal light reflex in LASIK for hyperopia. J Refract Surg. 2003;19:464–465.
    1. Arbelaez MC, Vidal C, Arba-Mosquera S. Clinical outcomes of corneal vertex versus central pupil references with aberration-free ablation strategies and LASIK. Invest Ophthalmol Vis Sci. 2008;49:5287–5294.
    1. Lee H, Kim TI, Kim EK. Corneal astigmatism analysis for toric intraocular lens implantation: precise measurements for perfect correction. Curr Opin Ophthalmol. 2015;26:34–38.
    1. Saad E, Shammas MC, Shammas HJ. Scheimpflug corneal power measurements for intraocular lens power calculation in cataract surgery. Am J Ophthalmol. 2013;156:460–467.
    1. Wegener A, Laser-Junga H. Photography of the anterior eye segment according to Scheimpflug's principle: options and limitations: a review. Clin Exp Ophthalmol. 2009;37:144–154.

Publication types

LinkOut - more resources