Larger and deeper ventricular lesions using a novel expandable spherical monopolar irrigated radiofrequency ablation catheter
- PMID: 31347223
- DOI: 10.1111/jce.14089
Larger and deeper ventricular lesions using a novel expandable spherical monopolar irrigated radiofrequency ablation catheter
Abstract
Background: Radiofrequency (RF) ablation is an established treatment for ventricular tachycardia (VT). However, the inability of current RF catheters to address deep or large substrate may explain most of the clinical failures.
Objectives: The aim of this study is to assess the efficacy and safety of ablation in the left ventricle (LV) in sheep using a novel 8-Fr deflectable ablation catheter (Sphere-9; Affera, Inc) with a 9-mm expandable spherical monopolar irrigated RF tip vs a standard RF irrigated catheter (Biosense Webster, Diamond Bar, CA). The impact on tissue was assessed on local bipolar electrograms (from nine uniformly distributed mini surface electrodes and an internal central reference electrode), as well as on direct lesion measurement post mortem.
Methods and results: Eleven sheep underwent LV endocardial ablation in healthy tissue using the Sphere-9 catheter (n = 6), or a conventional irrigated RF catheter (n = 5). Twenty lesions were created with the Sphere-9 (current limit: 2.7 A; temp. limit: 60°C; irrigation: 30 mL/min; and duration: 60-120 seconds). Local bipolar electrograms at the surface of the catheter disappeared during RF delivery in 17 of 20 (85%) lesions. The mean lesion volume was 1707 ± 771 mm 3 (length: 15.8 ± 3.3 mm; width: 11.6 ± 4.2 mm; and depth: 10.3 ± 2.9 mm). Twenty-five lesions were created with a standard RF irrigated catheter (power control 35 W; irrigation: 30 mL/min; duration: 60 seconds; volume 537 ± 398 mm 3 ; length: 8.2 ± 2.3 mm; width: 5.2 ± 1.8 mm; and depth: 5.5 ± 2.4 mm). The novel spherical RF catheter created significantly larger lesions ( P < .001 for measurements in all dimensions). There were no steam pops with the novel ablation catheter vs one with the conventional catheter.
Conclusions: This novel spherical monopolar irrigated RF catheter creates lesions that are twice as large and deep as a standard irrigated RF catheter.
Keywords: ablation; novel ablation catheter; spherical monopolar irrigated ablation catheter; ventricular tachycardia.
© 2019 Wiley Periodicals, Inc.
Similar articles
-
Comparison of ventricular radiofrequency lesions in sheep using standard irrigated tip catheter versus catheter ablation enabling direct visualization.J Cardiovasc Electrophysiol. 2012 Aug;23(8):869-73. doi: 10.1111/j.1540-8167.2012.02338.x. Epub 2012 May 3. J Cardiovasc Electrophysiol. 2012. PMID: 22554117
-
Irrigated Needle Ablation Creates Larger and More Transmural Ventricular Lesions Compared With Standard Unipolar Ablation in an Ovine Model.Circ Arrhythm Electrophysiol. 2015 Dec;8(6):1498-506. doi: 10.1161/CIRCEP.115.002963. Epub 2015 Sep 10. Circ Arrhythm Electrophysiol. 2015. PMID: 26359480
-
Effect of Irrigant Characteristics on Lesion Formation After Radiofrequency Energy Delivery Using Ablation Catheters with Actively Cooled Tips.J Cardiovasc Electrophysiol. 2015 Jul;26(7):792-8. doi: 10.1111/jce.12682. J Cardiovasc Electrophysiol. 2015. PMID: 25864402
-
Biophysics and clinical utility of irrigated-tip radiofrequency catheter ablation.Expert Rev Med Devices. 2012 Jan;9(1):59-70. doi: 10.1586/erd.11.42. Expert Rev Med Devices. 2012. PMID: 22145841 Review.
-
Forging ahead: Update on radiofrequency ablation technology and techniques.J Cardiovasc Electrophysiol. 2020 Jan;31(1):360-369. doi: 10.1111/jce.14317. Epub 2019 Dec 18. J Cardiovasc Electrophysiol. 2020. PMID: 31828880 Review.
Cited by
-
Exploring the Full Potential of Radiofrequency Technology: A Practical Guide to Advanced Radiofrequency Ablation for Complex Ventricular Arrhythmias.Curr Cardiol Rep. 2024 May;26(5):269-282. doi: 10.1007/s11886-024-02048-z. Epub 2024 May 3. Curr Cardiol Rep. 2024. PMID: 38700597 Free PMC article. Review.
-
Mapping and ablation of ventricular tachycardia using dual-energy lattice-tip focal catheter: early feasibility and safety study.Europace. 2024 Nov 1;26(11):euae275. doi: 10.1093/europace/euae275. Europace. 2024. PMID: 39478679 Free PMC article.
-
Novel Ablation Catheters for Atrial Fibrillation.Rev Cardiovasc Med. 2024 May 23;25(5):187. doi: 10.31083/j.rcm2505187. eCollection 2024 May. Rev Cardiovasc Med. 2024. PMID: 39076471 Free PMC article. Review.
-
De novo ventricular tachycardia ablation with stacked pulsed-field and radiofrequency energy via a dual-energy lattice-tipped catheter.HeartRhythm Case Rep. 2025 Jan 31;11(5):389-394. doi: 10.1016/j.hrcr.2025.01.016. eCollection 2025 May. HeartRhythm Case Rep. 2025. PMID: 40656004 Free PMC article. No abstract available.