Observational Study
. 2019 Aug;35(8):539-543.
doi: 10.1097/PEC.0000000000001235.

Efficacy of Rapid Fluid Administration Using Various Setups and Devices

Affiliations
Observational Study

Efficacy of Rapid Fluid Administration Using Various Setups and Devices

Onur Balaban et al. Pediatr Emerg Care. 2019 Aug.

Abstract

Objectives: In clinical practice, there are various methods that can be used for the rapid administration of fluid in infants and children. The current study prospectively evaluates gravity, pressure-assisted, and hand-pump methods for the rapid administration of fluid using an in vitro model.

Methods: Thirty participants were asked to deliver 500 mL of fluid using 1 of 6 setups: (1) standard blood tubing with gravity administration, (2) standard blood tubing with pressure bag maintained at 300 mm Hg, (3) standard blood tubing with pressure bag inflated to 300 mm Hg and left to flow, (4) blood tubing with in-line bulb pump, (5) blood tubing with in-line bulb pump and pressure bag, and (6) standard blood tubing with 20-mL syringe attached to the stopcock for a push-and-pull technique using a 20-mL syringe.

Results: The blood tubing with an in-line bulb pump to allow manual acceleration of the administration of fluid along with a pressure bag on the intravenous fluid bag achieved the fastest flow rate, requiring an average of 98 seconds to deliver 500 mL of fluid.

Conclusions: When considering factors that affect fluid administration, Poiseuille's law dictates that the most important variable is the radius of the intravenous cannula, whereas the length of the cannula and the viscosity of the fluid administered are of secondary importance. With these limitations in mind, other factors may be used to speed fluid administration. Our study demonstrates the advantage of using blood tubing with the in-line bulb pump combined with a pressure bag.