Solid-State Detector SPECT Myocardial Perfusion Imaging
- PMID: 31375568
- DOI: 10.2967/jnumed.118.220657
Solid-State Detector SPECT Myocardial Perfusion Imaging
Abstract
There has been an evolutionary leap in SPECT imaging with the advent of camera systems that use solid-state crystals and novel collimator designs configured specifically for cardiac imaging. Solid-state SPECT camera systems have facilitated dramatic reductions in both imaging time and radiation dose while maintaining high diagnostic accuracy. These advances are related to simultaneous improvement in photon sensitivity due to the collimator and imaging geometry, as well as image resolution due to the improved energy resolution of the new crystals. Improved photon sensitivity has facilitated fast or low-dose myocardial perfusion imaging (MPI), and early dynamic imaging has emerged as a technique for assessing myocardial blood flow with SPECT. Lastly, general-purpose solid-state camera systems and hybrid SPECT/CT systems have also been developed that may have important clinical roles in cardiac imaging. This review summarizes state-of-the-art solid-state SPECT MPI technology and clinical applications, including emerging techniques for SPECT MPI flow estimation. We also discuss imaging protocols used with the new cameras, potential imaging pitfalls, and the latest data providing large-scale validation of the diagnostic and prognostic value of this new technology.
Keywords: SPECT MPI; instrumentation; solid-state SPECT.
© 2019 by the Society of Nuclear Medicine and Molecular Imaging.
Similar articles
-
High-speed myocardial perfusion imaging initial clinical comparison with conventional dual detector anger camera imaging.JACC Cardiovasc Imaging. 2008 Mar;1(2):156-63. doi: 10.1016/j.jcmg.2007.12.004. JACC Cardiovasc Imaging. 2008. PMID: 19356422
-
Additional diagnostic value of integrated analysis of cardiac CTA and SPECT MPI using the SMARTVis system in patients with suspected coronary artery disease.J Nucl Med. 2014 Jan;55(1):50-7. doi: 10.2967/jnumed.113.119842. Epub 2013 Dec 12. J Nucl Med. 2014. PMID: 24337600
-
Approaches to reducing radiation dose from radionuclide myocardial perfusion imaging.J Nucl Med. 2015 Apr;56(4):592-9. doi: 10.2967/jnumed.112.115097. Epub 2015 Mar 12. J Nucl Med. 2015. PMID: 25766891 Review.
-
Real-time respiratory triggered SPECT myocardial perfusion imaging using CZT technology: impact of respiratory phase matching between SPECT and low-dose CT for attenuation correction.Eur Heart J Cardiovasc Imaging. 2017 Jan;18(1):31-38. doi: 10.1093/ehjci/jew031. Epub 2016 Mar 16. Eur Heart J Cardiovasc Imaging. 2017. PMID: 26985074
-
New solid state cadmium-zinc-telluride technology for cardiac single photon emission computed tomographic myocardial perfusion imaging.Expert Rev Med Devices. 2017 Mar;14(3):213-222. doi: 10.1080/17434440.2017.1296763. Epub 2017 Mar 1. Expert Rev Med Devices. 2017. PMID: 28276752 Review.
Cited by
-
Feasibility of Imaging Small Animals on a 360° Whole-Body Cadmium Zinc Telluride SPECT Camera: a Phantom Study.Mol Imaging Biol. 2022 Dec;24(6):1018-1027. doi: 10.1007/s11307-022-01753-x. Epub 2022 Jul 14. Mol Imaging Biol. 2022. PMID: 35835951
-
Monte Carlo Simulation and Reconstruction: Assessment of Myocardial Perfusion Imaging of Tracer Dynamics With Cardiac Motion Due to Deformation and Respiration Using Gamma Camera With Continuous Acquisition.Front Cardiovasc Med. 2022 Jul 13;9:871967. doi: 10.3389/fcvm.2022.871967. eCollection 2022. Front Cardiovasc Med. 2022. PMID: 35911544 Free PMC article.
-
Cumulative radiation doses due to nuclear medicine examinations: a systematic review.Br J Radiol. 2021 Oct 1;94(1126):20210444. doi: 10.1259/bjr.20210444. Epub 2021 Aug 11. Br J Radiol. 2021. PMID: 34379454 Free PMC article.
-
The effects of airway disease on the deposition of inhaled drugs.Expert Opin Drug Deliv. 2024 Aug;21(8):1175-1190. doi: 10.1080/17425247.2024.2392790. Epub 2024 Aug 19. Expert Opin Drug Deliv. 2024. PMID: 39136493 Free PMC article. Review.
-
Radiomics for the detection of diffusely impaired myocardial perfusion: A proof-of-concept study using 13N-ammonia positron emission tomography.J Nucl Cardiol. 2023 Aug;30(4):1474-1483. doi: 10.1007/s12350-022-03179-y. Epub 2023 Jan 5. J Nucl Cardiol. 2023. PMID: 36600174 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical