Randomized Controlled Trial
. 2019 Dec:208:219-225.
doi: 10.1016/j.ajo.2019.07.015. Epub 2019 Aug 1.

Safety and Effectiveness of CyPass Supraciliary Micro-Stent in Primary Open-Angle Glaucoma: 5-Year Results from the COMPASS XT Study

Affiliations
Randomized Controlled Trial

Safety and Effectiveness of CyPass Supraciliary Micro-Stent in Primary Open-Angle Glaucoma: 5-Year Results from the COMPASS XT Study

George Reiss et al. Am J Ophthalmol. 2019 Dec.

Abstract

Purpose: To characterize the long-term (up to 5 years) safety and effectiveness of the supraciliary Micro-Stent (Alcon) implanted at the time of phacoemulsification in eyes with coexisting open-angle glaucoma (OAG) and visually significant cataract.

Design: Three-year safety extension of a 2-year randomized clinical trial.

Methods: Patients from the multicenter Study of an Implantable Device for Lowering Intraocular Pressure in Glaucoma Patients Undergoing Cataract Surgery (COMPASS) trial who underwent Micro-Stent implantation plus phacoemulsification (n = 215) or phacoemulsification alone (n = 67) were evaluated 36, 48, and 60 months postoperatively. The primary outcome measurement was the occurrence of sight-threatening ocular adverse events. Evaluations at each time point included best-corrected visual acuity (BCVA), anterior and posterior segment examinations, tonometry, gonioscopy, pachymetry, perimetry, specular microscopy, and assessment of adverse events.

Results: Three sight-threatening ocular adverse events occurred, 2 in the Micro-Stent group and 1 in the control group, but none of these events was related to the Micro-Stent device. Ocular adverse events were of similar frequencies in both groups, the most common of which were BCVA loss of ≥2 lines compared with best BCVA in COMPASS and worsening of visual field mean defect (VFMD) ≥2.5 dB compared with month 24. Changes in mean BCVA from baseline, clinical examinations, pachymetry, and VFMD were similar in the 2 groups. At 60 months, a higher proportion of subjects in the Micro-Stent group (46%; 95% confidence interval [CI], 38.9%-53.2%) than in the control group (32.1%; 95% CI 19.9%-46.3%) were able to achieve a reduction of ≥20% in intraocular pressure without using hypotensive medication.

Conclusions: Few sight-threatening serious ocular adverse events occurred following Micro-Stent implantation, and clinical evidence of corneal decompensation was minimal.