Accuracy of the robot-assisted procedure in deep brain stimulation
- PMID: 31400032
- DOI: 10.1002/rcs.2032
Accuracy of the robot-assisted procedure in deep brain stimulation
Abstract
Introduction: The use of a robot-assisted technology becomes very competitive. The aim of this work was to define the accuracy of robotic assistance in deep brain stimulation surgery and to compare results with that in the literature.
Methods: We retrospectively reviewed the accuracy of lead implantation in 24 consecutive patients who had robot-assisted (ROSA, Zimmer-Biomet) surgery for the treatment of movement disorders. Intended stereotactic coordinates (x, y, z) of contact 0 (the most distal contact at the tip of the electrode) of each definitive lead were compared with actual coordinates obtained by a postoperative CT scan. For each lead, the euclidian 3D distance between the actual and intended location of contact 0 was calculated.
Results: The euclidian 3D distances between the intended and actual location of the contact 0 were 0.81 mm on the right side and 1.12 mm on the left side.
Discussion: Robot-assisted technology for stereotactic surgery is safe and accurate. The association with a 3D flat-panel CT scan is an optimized procedure for deep intracranial electrode implantation.
Keywords: accuracy; deep brain stimulation; movement disorder; robot-assisted surgery.
© 2019 John Wiley & Sons, Ltd.
References
REFERENCES
-
- Kalan S, Chauhan S, Coelho RF, et al. History of robotic surgery. J Robot Surg. 2010;4(3):141-147.
-
- Kwoh YS, Hou J, Jonckheere EA, Hayati S. A robot with improved absolute positioning accuracy for CT guided stereotactic brain surgery. IEEE Trans Biomed Eng. 1988;35(2):153-160.
-
- Derrey S, Maltête D, Chastan N, et al. Deep brain stimulation of the subthalamic nucleus in Parkinson's disease: usefulness of intraoperative radiological guidance. Stereotact Funct Neurosurg. 2008;86(6):351-358.
-
- Weise LM, Seifried C, Eibach S, et al. Correlation of active contact positions with the electrophysiological and anatomical subdivisions of the subthalamic nucleus in deep brain stimulation. Stereotact Funct Neurosurg. 2013;91(5):298-305.
-
- Longhi M, Ricciardi G, Tommasi G, et al. The role of 3T magnetic resonance imaging for targeting the human subthalamic nucleus in deep brain stimulation for Parkinson disease. J Neurol Surgery Part A: Cen Eur Neurosurg. 2015;76(03):181-189.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical