Efficacy of two vitamin D immunoassays to detect 25-OH vitamin D2 and D3

Affiliations

Efficacy of two vitamin D immunoassays to detect 25-OH vitamin D2 and D3

Emily Garnett et al. Pract Lab Med. .

Abstract

Objectives: Assessment of Vitamin D status by measurement of 25-Hydroxyvitamin D (25-OH-D) is widely performed by immunoassay. Yet, the ability of these assays to detect Vitamin D2 (as 25-OH-D2) or Vitamin D3 (as 25-OH-D3) varies. It is important to recognize the ability of an assay to quantitate either form of 25-OH-D to evaluate Vitamin D status of supplemented patients. We evaluated detection of 25-OH-D2 and 25-OH-D3 by two assays in our medical center.

Design and methods: The Abbott Architect i1000 SR 25-OH Vitamin D assay and Roche Cobas 8000 Vitamin D assay were compared for their recovery of 25-OH-D2 or D3 from spiked serum samples. Samples with known endogenous concentrations of 25-OH-D2 or D3 by LC-MS/MS were also measured to calculate bias between our assays and LC-MS/MS.

Results: Recovery of 25-OH-D3 in spiked samples was similar by Architect (84-87%) and Cobas (90%). Recovery of 25-OH-D2 was lower than 25-OH-D3, and was poorer by Architect (37-40%) than by Cobas (69-71%). In measurement of samples with known 25-OH-D concentrations, performance of Architect and Cobas assays was similar for 25-OH-D3. However, at concentrations >50 nmol/L 25-OH-D2, the Architect assay exhibited large average negative bias (-27%).

Conclusions: While the Architect and Cobas assays performed similarly in detection of 25-OH-D3, the Architect assay was significantly poorer at detecting 25-OH-D2 than Cobas, with poorer recovery and significant negative bias at higher concentrations of 25-OH-D2. This agrees with other studies, and indicates that caution should be used in interpreting Architect 25-OH-D results in patients supplemented with Vitamin D2.

Keywords: 25-Hydroxyvitamin D; Assay comparison; Immunoassay; Recovery; Vitamin D2.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
Bias plots of total 25-OH-D assay results relative to 25-OH-D2 or D3 by LC-MS/MS. Samples were measured in singlicate. Percent bias was calculated for Architect and Cobas assays for measurements of samples with A) known 25-OH-D3 or B) known 25-OH-D2 concentrations.

Similar articles

Cited by

References

    1. Wyness S.P., Straseski J.A. Performance characteristics of six automated 25-hydroxyvitamin D assays: mind your 3s and 2s. Clin. Biochem. 2015;48:1089–1096. 2015/08/15. - PubMed
    1. Freeman J., Wilson K., Spears R. Performance evaluation of four 25-hydroxyvitamin D assays to measure 25-hydroxyvitamin D2. Clin. Biochem. 2015;48:1097–1104. 2015/06/10. - PubMed
    1. Garg U. 25-Hydroxyvitamin D testing: immunoassays versus tandem mass spectrometry. Clin. Lab. Med. 2018;38:439–453. 2018/08/18. - PubMed
    1. Abbott ARCHITECT 25-OH Vitamin D Assay [directional Insert 3L52, G3-3094/R05] (Package Insert) Abbott Laboratories; Abbott Park, IL, USA: 2013.
    1. Roche Cobas 25-Hydroxyvitamin D Assay [directional Insert 2015-05, V 6.0 English] (Package Insert) Roche Diagnostics; Indianapolis, IN, USA: 2015.