Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Clinical Trial
. 2020 Jan-Feb;10(1):e37-e44.
doi: 10.1016/j.prro.2019.08.009. Epub 2019 Aug 31.

Anchored Transponder Guided Lung Radiation Therapy

Affiliations
Clinical Trial

Anchored Transponder Guided Lung Radiation Therapy

Michael C Dobelbower et al. Pract Radiat Oncol. 2020 Jan-Feb.

Abstract

Purpose: The Calypso Beacon transponder has been modified by the addition of a nitinol anchor feature to allow for positional stability when implanted bronchoscopically into the lung. The purpose of this study was to confirm the feasibility and safety of anchored transponder placement and feasibility of lung target localization and tracking.

Methods and materials: This study enrolled patients with histologically confirmed cancer in the lung (primary or metastatic) who were scheduled to receive external beam radiation therapy. Three anchored transponders were implanted via flexible bronchoscopy into small (approximately 2- to 2.5-mm diameter) airways. Patient alignment at each radiation fraction was performed with the Calypso system, and anchored transponder position was tracked during radiation delivery. The primary endpoint was defined as the ability to localize at least 85% of the patients during the first week of treatment. Four follow-up visits were specified including a posttreatment assessment and every 3 months up to 1 year.

Results: A total of 69 patients underwent anchored transponder placement, and all 207 implanted anchored transponders were visible on the treatment-planning simulation computed tomography scan. Sixty-seven patients underwent radiation therapy, and localization was successful in 66 cases (98.5%). With 1 failure in 67 cases, the P value for rejecting the null hypothesis was <.001 and the primary objective of the study met. Five adverse events in 5 patients were potentially attributed to the study device or implantation procedure, consisting of pneumonia (2 cases), pleural abscess (1 case), and pneumothorax (2 cases). Two serious events (cardiac arrest and acute hypotension) were attributed to anesthesia during the implantation procedure.

Conclusions: This study strongly supports that anchored transponders are safe, positionally stable, and useful for lung tumor localization and monitoring.