. 2019 Nov;257(11):2437-2447.
doi: 10.1007/s00417-019-04453-5. Epub 2019 Sep 3.

Validation of corneal topographic and aberrometric measurements obtained by color light-emitting diode reflection topography in healthy eyes

Affiliations

Validation of corneal topographic and aberrometric measurements obtained by color light-emitting diode reflection topography in healthy eyes

David P Piñero et al. Graefes Arch Clin Exp Ophthalmol. 2019 Nov.

Abstract

Purpose: To evaluate the intrasession repeatability of anterior corneal topographic and aberrometric measurements provided by a color-LED topographer as well as their interchangeability with those provided by a Scheimpflug-based system in healthy eyes.

Methods: Thirty-five healthy eyes of 35 patients (age, 16-66 years) were enrolled. A complete eye examination was performed in all cases including a complete corneal analysis with the Scheimpflug-based system Pentacam (Oculus Optikgeräte) (one measurement) and the Cassini system (i-Optics) (three consecutive measurements). Intrasession repeatability of the Cassini measurements was assessed with the within-subject standard deviation (Sw) and the intraclass correlation coefficient (ICC). The Bland-Altman analysis was used to evaluate the agreement between both devices.

Results: Mean Sw for keratometric readings was 0.02 mm (ICC ≥ 0.992), ranging between 0.16 and 0.05 D (ICC 0.930-0.978) for anterior and total astigmatic measurements. Mean Sw for asphericity and corneal diameter were 0.06 (ICC 0.926) and 0.03 mm (IC 0.997), respectively. Aberrometric parameters showed ICCs ≥ 0.816, except for Z42 (ICC 0.741) and Z44 (ICC 0.544). When comparing devices, statistically significant differences were found for most of topographic and aberrometric data (p ≤ 0.044). Likewise, ranges of agreement between devices were clinically relevant (keratometry > 0.06 mm; total astigmatic components > 0.69 D; asphericity 0.35; second-, third-, and fourth-order Zernike terms, more than 0.20, 0.13, and 0.01 μm, respectively).

Conclusions: Consistent anterior corneal topographic, total corneal astigmatic, and aberrometric measurements are obtained with color-LED topography in healthy eyes, which are not interchangeable with those provided by the Scheimpflug-based topography.

Keywords: Aberrometry; Anterior corneal curvature; Cassini; Color light-emitting diode reflection topography; Corneal topography; Pentacam; Posterior corneal curvature.

PubMed Disclaimer

Similar articles

Cited by

References

    1. J Cataract Refract Surg. 2015 Jan;41(1):186-92 - PubMed
    1. Clin Ophthalmol. 2015 Feb 05;9:245-52 - PubMed
    1. J Cataract Refract Surg. 2014 Feb;40(2):269-75 - PubMed
    1. J Cataract Refract Surg. 2008 May;34(5):727-34 - PubMed
    1. Am J Ophthalmol. 2016 Jul;167:72-8 - PubMed

Publication types

LinkOut - more resources