Reducing preterm mortality in eastern Uganda: the impact of introducing low-cost bubble CPAP on neonates <1500 g
- PMID: 31484567
- PMCID: PMC6724283
- DOI: 10.1186/s12887-019-1698-x
Reducing preterm mortality in eastern Uganda: the impact of introducing low-cost bubble CPAP on neonates <1500 g
Abstract
Background: Complications of prematurity are the leading cause of deaths in children under the age of five. The predominant reason for these preterm deaths is respiratory distress syndrome (RDS). In low-income countries (LICs) there are limited treatment options for RDS. Due to their simplicity and affordability, low-cost bubble continuous positive airway pressure (bCPAP) devices have been introduced in neonatal units in LICs to treat RDS. This study is the first observational study from a LIC to compare outcomes of very-low-birth-weight (VLBW) neonates in pre- and post-CPAP periods.
Methods: This was a retrospective study of VLBW neonates (weight < 1500 g) in Mbale Regional Referral Hospital Neonatal Unit (MRRH-NNU), a government hospital in eastern Uganda. It aimed to measure the outcome of VLBW neonates in two distinct study periods: A 14-month period beginning at the opening of MRRH-NNU and covering the period until bCPAP was introduced (pre-bCPAP) and an 18-month period following the introduction of bCPAP (post-bCPAP). After the introduction of bCPAP, it was applied to preterm neonates with RDS when clinically indicated and if a device was available. Clinical features and outcomes of all neonates < 1500 g were compared before and after the introduction of bCPAP.
Results: The admission records of 377 VLBW neonates < 1500 g were obtained. One hundred fifty-eight were admitted in the pre-bCPAP period and 219 in the post-bCPAP period. The mortality rate in the pre- bCPAP period was 39.2% (62/158) compared with 26.5% (58/219, P = 0.012) in the post-bCPAP period. Overall, there was a 44% reduction in mortality (OR 0.56, 95%CI 0.36-0.86, P = 0.01). There were no differences in birthweight, sex, presence of signs of respiratory distress or apnoea between the two groups.
Conclusion: Specialized and resource-appropriate neonatal care, that appropriately addresses the challenges of healthcare provision in LICs, has the potential to reduce neonatal deaths. The use of a low-cost bCPAP to treat RDS in VLBW neonates resulted in a significant improvement in their survival in a neonatal unit in eastern Uganda. Since RDS is one of the leading causes of neonatal mortality, it is possible that this relatively simple and affordable intervention could have a huge impact on global neonatal mortality.
Keywords: Africa; Bubble CPAP; Low-income country; Neonate; Preterm; Respiratory distress syndrome.
Conflict of interest statement
The authors declare that they have no competing interests.
Figures
References
-
- Liu L, Oza S, Hogan D, Chu Y, Perin J, Zhu J, et al. Global, regional, and national causes of under-5 mortality in 2000-15: an updated systematic analysis with implications for the sustainable development goals. Lancet. 2016;388(10063):3027–3035. doi: 10.1016/S0140-6736(16)31593-8. - DOI - PMC - PubMed