Dual-Energy Low-keV or Single-Energy Low-kV CT for Endoleak Detection?: A 6-Reader Study in an Aortic Aneurysm Phantom
- PMID: 31503078
- DOI: 10.1097/RLI.0000000000000606
Dual-Energy Low-keV or Single-Energy Low-kV CT for Endoleak Detection?: A 6-Reader Study in an Aortic Aneurysm Phantom
Abstract
Objectives: The aim of this study was to compare image quality, conspicuity, and endoleak detection between single-energy low-kV images (SEIs) and dual-energy low-keV virtual monoenergetic images (VMIs+) in computed tomography angiography of the aorta after endovascular repair.
Materials and methods: An abdominal aortic aneurysm phantom simulating 36 endoleaks (2 densities; diameters: 2, 4, and 6 mm) in a medium- and large-sized patient was used. Each size was scanned using single-energy at 80 kVp (A) and 100 kVp (B), and dual-energy at 80/Sn150kVp for the medium (C) and 90/Sn150kVp for the large size (D). VMIs+ at 40 keV and 50 keV were reconstructed from protocols C and D. Radiation dose was 3 mGy for the medium and 6 mGy for the large size. Objective image quality and normalized noise power spectrum were determined. Subjective image quality, conspicuity, and sensitivity for endoleaks were independently assessed by 6 radiologists. Sensitivity was compared using Marascuilo procedure and Fisher exact test. Conspicuities were compared using Wilcoxon-matched pairs test, analysis of variance, and Tukey test.
Results: The contrast-to-noise-ratio of the aorta was significantly higher for VMI+ compared with SEI (P < 0.001). Noise power spectrum showed a higher noise magnitude and coarser texture in VMI+. Subjective image quality and overall conspicuity was lower for VMI+ compared with SEI (P < 0.05). Sensitivity for endoleaks was overall higher in the medium phantom for SEI (60.9% for A, 62.2% for B) compared with VMI+ (54.2% for C, 49.3% for D) with significant differences between protocols B and D (P < 0.05). In the large phantom, there was no significant difference in sensitivity among protocols (P = 0.79), with highest rates for protocols B (31.4%) and C (31.7%).
Conclusions: Our study indicates that low-keV VMI+ results in improved contrast-to-noise-ratio of the aorta, whereas noise properties, subjective image quality, conspicuity, and sensitivity for endoleaks were overall superior for SEI.
Similar articles
-
Enhanced visualization in endoleak detection through iterative and AI-noise optimized spectral reconstructions.Sci Rep. 2024 Feb 15;14(1):3845. doi: 10.1038/s41598-024-54502-1. Sci Rep. 2024. PMID: 38360941 Free PMC article.
-
Multi-Energy Low-Kiloelectron Volt versus Single-Energy Low-Kilovolt Images for Endoleak Detection at CT Angiography of the Aorta.Radiol Cardiothorac Imaging. 2024 Apr;6(2):e230217. doi: 10.1148/ryct.230217. Radiol Cardiothorac Imaging. 2024. PMID: 38451189 Free PMC article. Clinical Trial.
-
Endoleaks after endovascular aortic aneurysm repair: Improved detection with noise-optimized virtual monoenergetic dual-energy CT.Eur J Radiol. 2017 Sep;94:125-132. doi: 10.1016/j.ejrad.2017.06.017. Epub 2017 Jun 27. Eur J Radiol. 2017. PMID: 28712695
-
Prospective Comparison of 70-kVp Single-Energy CT versus Dual-Energy CT: Which is More Suitable for CT Angiography with Low Contrast Media Dosage?Acad Radiol. 2020 May;27(5):e116-e122. doi: 10.1016/j.acra.2019.07.016. Epub 2019 Sep 16. Acad Radiol. 2020. PMID: 31537504 Clinical Trial.
-
Noise-optimized virtual monoenergetic imaging technology of the third-generation dual-source computed tomography and its clinical applications.Quant Imaging Med Surg. 2021 Nov;11(11):4627-4643. doi: 10.21037/qims-20-1196. Quant Imaging Med Surg. 2021. PMID: 34737929 Free PMC article. Review.
Cited by
-
Enhanced visualization in endoleak detection through iterative and AI-noise optimized spectral reconstructions.Sci Rep. 2024 Feb 15;14(1):3845. doi: 10.1038/s41598-024-54502-1. Sci Rep. 2024. PMID: 38360941 Free PMC article.
-
Multi-Energy Low-Kiloelectron Volt versus Single-Energy Low-Kilovolt Images for Endoleak Detection at CT Angiography of the Aorta.Radiol Cardiothorac Imaging. 2024 Apr;6(2):e230217. doi: 10.1148/ryct.230217. Radiol Cardiothorac Imaging. 2024. PMID: 38451189 Free PMC article. Clinical Trial.
-
Review of Clinical Applications of Dual-Energy CT in Patients after Endovascular Aortic Repair.J Clin Med. 2023 Dec 18;12(24):7766. doi: 10.3390/jcm12247766. J Clin Med. 2023. PMID: 38137834 Free PMC article. Review.
-
Virtual monochromatic images of dual-energy CT as an alternative to single-energy CT: performance comparison using a detectability index for different acquisition techniques.Eur Radiol. 2023 Aug;33(8):5752-5760. doi: 10.1007/s00330-023-09491-6. Epub 2023 Mar 9. Eur Radiol. 2023. PMID: 36892640
-
Unenhanced computed tomography radiomics help detect endoleaks after endovascular repair of abdominal aortic aneurysm.Eur Radiol. 2024 Mar;34(3):1647-1658. doi: 10.1007/s00330-023-10000-y. Epub 2023 Sep 2. Eur Radiol. 2024. PMID: 37658886 Free PMC article.
References
-
- Bulder RMA, Bastiaannet E, Hamming JF, et al. Meta-analysis of long-term survival after elective endovascular or open repair of abdominal aortic aneurysm. Br J Surg. 2019;106:523–533.
-
- Cassagnes L, Perignon R, Amokrane F, et al. Aortic stent-grafts: endoleak surveillance. Diagn Interv Imaging. 2016;97:19–27.
-
- Guo Q, Zhao J, Huang B, et al. A systematic review of ultrasound or magnetic resonance imaging compared with computed tomography for endoleak detection and aneurysm diameter measurement after endovascular aneurysm repair. J Endovasc Ther. 2016;23:936–943.
-
- Javor D, Wressnegger A, Unterhumer S, et al. Endoleak detection using single-acquisition split-bolus dual-energy computer tomography (DECT). Eur Radiol. 2017;27:1622–1630.
-
- Partovi S, Trischman T, Rafailidis V, et al. Multimodality imaging assessment of endoleaks post-endovascular aortic repair. Br J Radiol. 2018;91:20180013.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical