Case Reports
. 2019 Sep-Oct;10(5):969-971.
doi: 10.1016/j.jcot.2019.01.014. Epub 2019 Jan 15.

Technical tip for removal of broken cephalomedullary reamer tip in cases of proximal femoral nailing

Affiliations
Case Reports

Technical tip for removal of broken cephalomedullary reamer tip in cases of proximal femoral nailing

Arvind Kumar et al. J Clin Orthop Trauma. 2019 Sep-Oct.

Abstract

Instrumentation breakage around hip joint can pose a challenging situation considering its vicinity to several vital structures. Broken fragments carry the risk of migration and thus should be removed as early as possible. A case of successful retrieval of broken tip of cephalomedullary lag screw reamer of a cephalomedullary nail, in basicervical region of femoral neck, during fixation of a subtrochanteric femoral fracture has been reported. Literature review has been done to suggest techniques to tackle similar situations using simple and commonly available instruments.

Keywords: Broken; Cephalomedullary; Hip; Instruments; Nail; Reamer.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
Intraoperative image intensifier pictures showing broken cephalomedullary lag screw reamer tip fragments.
Fig. 2
Fig. 2
Intraoperative image intensifier pictures showing the inferior fragment being scooped out towards the nail entry point using a curette.
Fig. 3
Fig. 3
The broken cephalomedullary lag screw reamer along retrieved broken fragments.

Similar articles

References

    1. Sen R.K., Tripathy S.K., Aggarwal S. Broken Kirschner or guide-wire retrieval: a report of 4 cases. Hip Int. 2010;20(4):551–554. - PubMed
    1. Sayegh F.E., Tsintzas D., Kapetanos G.A. Intrapelvic migration of a guide pin during fixation of a hip fracture: who and what is to blame? Acta Orthop Belg. 2005;71(2):239–241. - PubMed
    1. Riley P.M., Weiner D.S., Gillespie R., Weiner S.D. Hazards of internal fixation in the treatment of slipped capital femoral epiphysis. J Bone Joint Surg Am. 1990;72(10):1500–1509. - PubMed
    1. Afshar A. Intrapelvic protrusion of a broken guide wire fragment during fixation of a femoral neck fracture. Arch Bone Joint Surg. 2017;5(1):63–65. - PMC - PubMed
    1. Ilizaliturri V.M., Jr., Zarate-Kalfopulos B., Martinez-Escalante F.A., Cuevas-Olivo R., Camacho-Galindo J. Arthroscopic retrieval of a broken guide wire fragment from the hip joint alter cannulated screw fixation of slipped capital femoral epiphysis. Arthroscopy. 2007;23(2):227. e1–4. - PubMed

Publication types