Comparative Study
. 2021 Mar;105(3):440-444.
doi: 10.1136/bjophthalmol-2019-314031. Epub 2019 Sep 17.

Low-cost, smartphone-based frequency doubling technology visual field testing using a head-mounted display

Affiliations
Comparative Study

Low-cost, smartphone-based frequency doubling technology visual field testing using a head-mounted display

Karam AlRahman Alawa et al. Br J Ophthalmol. 2021 Mar.

Abstract

Background: Current visual field screening machines are bulky and expensive, limiting their accessibility, affordability and use. We report the design and evaluation of a novel, portable, cost-effective system for glaucoma screening using smartphone-based visual field screening using frequency doubling technology (FDT) and a head-mounted display.

Methods: Nineteen eyes of 10 subjects with new-onset or chronic primary open angle glaucoma were tested and compared with the Humphrey Zeiss FDT and the newly developed Mobile Virtual Perimetry (MVP) FDT with the C-20 testing pattern. Mann-Whitney, Bland-Altman and linear regression analyses were performed to assess statistical difference, agreement and correlation, respectively, between the two devices.

Results: The average age of the participants was 58±15 years. No statistically significant difference was found between the MVP FDT and the Humphrey Zeiss FDT (p>0.05). Bland-Altman and linear regression analyses demonstrated good agreement and correlation between the two devices.

Conclusion: The MVP FDT is a low-cost, portable visual field screening device that produces comparable results to the Humphrey Zeiss FDT and may be used as an easily accessible screening tool for glaucoma.

Keywords: Field of vision; Glaucoma; Telemedicine.