Comparative Study
. 2019 Oct;96(10):733-744.
doi: 10.1097/OPX.0000000000001435.

Validation of a Clinical Aberrometer Using Pyramidal Wavefront Sensing

Affiliations
Comparative Study

Validation of a Clinical Aberrometer Using Pyramidal Wavefront Sensing

Neeraj K Singh et al. Optom Vis Sci. 2019 Oct.

Abstract

Significance: Measurement of ocular aberrations is a critical component of many optical corrections.

Purpose: This study examines the accuracy and repeatability of a newly available high-resolution pyramidal wavefront sensor-based aberrometer (Osiris by Costruzione Strumenti Oftalmici, Firenze, Italy).

Methods: An engineered model eye and a dilated presbyopic eye were used to assess accuracy and repeatability of aberration measurements after systematic introduction of lower- and higher-order aberrations with calibrated trial lenses (sphere +10.00 to -10.00 D, and astigmatic -4.00 and -2.00 D with axis 180, 90, and 45°) and phase plates (-0.57 to 0.60 μm of Seidel spherical aberration defined over a 6-mm pupil diameter). Osiris aberration measurements were compared with those acquired on a previously calibrated COAS-HD aberrometer for foveal and peripheral optics both with and without multizone dual-focus contact lenses. The impact of simulated axial and lateral misalignment was evaluated.

Results: Root-mean-square errors for paraxial sphere (corneal plane), cylinder, and axis were, respectively, 0.07, 0.11 D, and 1.8° for the engineered model and 0.15, 0.26 D, and 2.7° for the presbyopic eye. Repeatability estimates (i.e., standard deviation of 10 repeat measures) for the model and presbyopic eyes were 0.026 and 0.039 D for spherical error. Root-mean-square errors of 0.01 and 0.02 μm, respectively, were observed for primary spherical aberration and horizontal coma (model eye). Foveal and peripheral measures of higher- and lower-order aberrations measured with the Osiris closely matched parallel data collected with the COAS-HD aberrometer both with and without dual-focus zonal bifocal contact lenses. Operator errors of focus and alignment introduced changes of 0.018 and 0.02 D/mm in sphere estimates.

Conclusions: The newly available clinical pyramidal aberrometer provided accurate and repeatable measures of lower- and higher-order aberrations, even in the challenging but clinically important cases of peripheral retina and multifocal optics.