Accuracy of digital impressions for implant-supported complete-arch prosthesis, using an auxiliary geometry part-An in vitro study
- PMID: 31610069
- DOI: 10.1111/clr.13549
Accuracy of digital impressions for implant-supported complete-arch prosthesis, using an auxiliary geometry part-An in vitro study
Abstract
Objectives: The aim of this study was to evaluate the accuracy of complete-arch digital impressions for fabrication of an implant-supported prosthesis in the edentulous maxilla using an auxiliary geometry part.
Materials and methods: A replica of the upper jaw of an edentulous patient with four scannable impression copings was fabricated in stainless steel. This model was scanned with an industrial non-contact 3D structured blue light 3D scanner, and the measurements of three reference distances were established as reference values. Subsequently, the model was scanned in two different scenarios (with or without an auxiliary geometry part put in place and fixed to the model) using three intraoral scanners. Measurements were taken with 3D inspection software, and a digital impression of the complete arch was built with mesh processing software by combining 2 STL files obtained with an intraoral scanner.
Results: All measurements with the auxiliary geometry part gave significantly more accurate results (p < .05). Trueness improved in the three reference distances, reaching values of 8 ± 6 µm at D12 reference distance, 20 ± 11 µm at D13 and 35 ± 22 µm at D14. Precision also improved significantly with the use of the auxiliary geometry part placed on the model (p < .05). The best precision results at reference distances D13 and D14 were obtained with the True Definition Scanner.
Conclusions: The proposed methodology significantly improves the accuracy of complete-arch digital impressions in edentulous patients obtained in vitro, regardless of which scanner is used.
Keywords: accuracy; complete arch; digital impression; edentulous; implant-supported framework; intraoral scan.
© 2019 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.
References
REFERENCES
-
- Ahrberg, D., Lauer, H. C., Ahrberg, M., & Weigl, P. (2016). Evaluation of fit and efficiency of CAD/CAM fabricated all-ceramic restorations based on direct and indirect digitalization: A double-blinded, randomized clinical trial. Clinical Oral Investigations, 20(2), 291-300. https://doi.org/10.1007/s00784-015-1504-6
-
- Amin, S., Weber, H. P., Finkelman, M., El Rafie, K., Kudara, Y., & Papaspyridakos, P. (2017). Digital vs. conventional full-arch implant impressions: A comparative study. Clinical Oral Implants Research, 28(11), 1360-1367. https://doi.org/10.1111/clr.12994
-
- Andriessen, F. S., Rijkens, D. R., Van Der Meer, W. J., & Wismeijer, D. W. (2014). Applicability and accuracy of an intraoral scanner for scanning multiple implants in edentulous mandibles: A pilot study. Journal of Prosthetic Dentistry, 111(3), 186-194. https://doi.org/10.1016/j.prosdent.2013.07.010
-
- Carr, A. B., & Toth, B. (1995). Fit discrimination of full arch implant frameworks. Journal of Dental Research, 74, 74-151.
-
- Christensen, G. J. (1966). Marginal fit of gold inlay castings. The Journal of Prosthetic Dentistry, 16(2), 297-305. https://doi.org/10.1016/0022-3913(66)90082-5