Clinical Trial
. 2019 Oct 22;74(16):2047-2055.
doi: 10.1016/j.jacc.2019.07.084.

Implanted Monitor Alerting to Reduce Treatment Delay in Patients With Acute Coronary Syndrome Events

Affiliations
Free article
Clinical Trial

Implanted Monitor Alerting to Reduce Treatment Delay in Patients With Acute Coronary Syndrome Events

David R Holmes Jr et al. J Am Coll Cardiol. .
Free article

Abstract

Background: Increased pre-hospital delay during acute coronary syndrome (ACS) events contributes to worse outcome.

Objectives: The purpose of this study was to assess the effectiveness of an implanted cardiac monitor with real-time alarms for abnormal ST-segment shifts to reduce pre-hospital delay during ACS events.

Methods: In the ALERTS (AngeLmed Early Recognition and Treatment of STEMI) pivotal study, subjects at high risk for recurrent ACS events (n = 907) were randomized to control (Alarms OFF) or treatment groups for 6 months, after which alarms were activated in all subjects (Alarms ON). Emergency department (ED) visits with standard-of-care cardiac test results were independently adjudicated as true- or false-positive ACS events. Alarm-to-door (A2D) and symptom-to-door (S2D) times were calculated for true-positive ACS ED visits triggered by 3 possible prompts: alarm only, alarms + symptoms, or symptoms only.

Results: The Alarms ON group showed reduced delays, with 55% (95% confidence interval [CI]: 46% to 63%) of ED visits for ACS events <2 h compared with 10% (95% CI: 2% to 27%) in the Alarms OFF group (p < 0.0001). Results were similar when restricted to myocardial infarction (MI) events. Median pre-hospital delay for MI was 12.7 h for Alarms OFF and 1.6 h in Alarms ON subjects (p < 0.0089). Median A2D delay was 1.4 h for asymptomatic MI. Median S2D delay for symptoms-only MI (no alarm) in Alarms ON was 4.3 h.

Conclusions: Intracardiac monitoring with real-time alarms for ST-segment shift that exceeds a subject's self-normative ischemia threshold level significantly reduced the proportion of pre-hospital delays >2 h for ACS events, including asymptomatic MI, compared with symptoms-only ED visits in Alarms OFF. (AngeLmed for Early Recognition and Treatment of STEMI [ALERTS]; NCT00781118).

Keywords: implantable cardiac monitor; ischemia monitoring; pre-hospital delay; silent myocardial infarction; supply-side ischemia; symptom-to-door time.

PubMed Disclaimer

Comment in

  • On "Alert" for Acute Coronary Syndromes.
    Domanski MJ. Domanski MJ. J Am Coll Cardiol. 2019 Oct 22;74(16):2056-2057. doi: 10.1016/j.jacc.2019.08.1011. J Am Coll Cardiol. 2019. PMID: 31623763 No abstract available.