. 2020 Apr;83(4):1390-1404.
doi: 10.1002/mrm.28004. Epub 2019 Oct 21.

Dehydration assessment via a portable, single sided magnetic resonance sensor

Affiliations

Dehydration assessment via a portable, single sided magnetic resonance sensor

Ashvin Bashyam et al. Magn Reson Med. 2020 Apr.

Abstract

Purpose: Undiagnosed dehydration compromises health outcomes across many populations. Existing dehydration diagnostics require invasive bodily fluid sampling or are easily confounded by fluid and electrolyte intake, environment, and physical activity limiting widespread adoption. We present a portable MR sensor designed to measure intramuscular fluid shifts to identify volume depletion.

Methods: Fluid loss is induced via a mouse model of thermal dehydration (37°C; 15-20% relative humidity). We demonstrate quantification of fluid loss induced by hyperosmotic dehydration with multicomponent T2 relaxometry using both a benchtop NMR system and MRI localized to skeletal muscle tissue. We then describe a miniaturized (~1000 cm3 ) portable (~4 kg) MR sensor (0.28 T) designed to identify dehydration-induced fluid loss. T2 relaxometry measurements were performed using a Carr-Purcell-Meiboom-Gill pulse sequence in ~4 min.

Results: T2 values from the portable MR sensor exhibited strong (R2 = 0.996) agreement with benchtop NMR spectrometer. Thermal dehydration induced weight loss of 4 to 11% over 5 to 10 h. Fluid loss induced by thermal dehydration was accurately identified via whole-animal NMR and skeletal muscle. The portable MR sensor accurately identified dehydration via multicomponent T2 relaxometry.

Conclusion: Performing multicomponent T2 relaxometry localized to the skeletal muscle with a miniaturized MR sensor provides a noninvasive, physiologically relevant measure of dehydration induced fluid loss in a mouse model. This approach offers sensor portability, reduced system complexity, fully automated operation, and low cost compared with MRI. This approach may serve as a versatile and portable point of care technique for dehydration monitoring.

Keywords: Unilateral Linear Halbach; dehydration; multicomponent T2 relaxometry; portable magnetic resonance; single-sided NMR; tissue fluid distribution.

PubMed Disclaimer

References

REFERENCES

    1. Cheuvront SN, Kenefick RW. Dehydration: physiology, assessment, and performance effects. Compr Physiol. 2014;4:257-285.
    1. Seymour DG, Henschke PJ, Cape RD, Campbell AJ. Acute confusional states and dementia in the elderly: the role of dehydration/volume depletion, physical illness and age. Age Ageing. 1980;9:137-146.
    1. Chan HYL, Cheng A, Cheung SSS, et al. Association between dehydration on admission and postoperative complications in older persons undergoing orthopaedic surgery. J Clin Nurs. 2018;27:3679-3686.
    1. Lavizzo-Mourey R, Johnson J, Stolley P. Risk factors for dehydration among elderly nursing home residents. J Am Geriatr Soc. 1988;36:213-218.
    1. Thomas DR, Cote TR, Lawhorne L, et al. Understanding clinical dehydration and its treatment. J Am Med Dir Assoc. 2008;9:292-301.