Results of a Multicenter Feasibility Study of an Automated Bedside Glucose Monitoring System in the Burn Intensive Care Setting

Affiliations
Observational Study

Results of a Multicenter Feasibility Study of an Automated Bedside Glucose Monitoring System in the Burn Intensive Care Setting

Craig T Elder et al. J Burn Care Res. .

Abstract

Intensive blood glucose regimens required for tight glycemic control in critically ill burn patients carry risk of hypoglycemia and are ultimately limited by the frequency of which serum glucose measurements can be feasibly monitored. Continuous inline glucose monitoring has the potential to significantly increase the frequency of serum glucose measurement. The objective of this study was to assess the accuracy of a continuous glucose monitor with inline capability (Optiscanner) in the burn intensive care setting. A multicenter, observational study was conducted at two academic burn centers. One hundred and six paired blood samples were collected from 10 patients and measured on the Optiscanner and the Yellow Springs Instrument. Values were plotted on a Clarke Error Grid and mean absolute relative difference calculated. Treatment was guided by existing hospital protocols using separately obtained values. 97.2% of results obtained from Optiscanner were within 25% of corresponding Yellow Springs Instrument values and 100% were within 30%. Mean absolute relative difference was calculated at 9.6%. Our findings suggest that a continuous glucose monitor with inline capability provides accurate blood glucose measurements among critically ill burn patients.

PubMed Disclaimer

Publication types