Reverse Total Shoulder Arthroplasty: Implant Design Considerations

Affiliations
Review

Reverse Total Shoulder Arthroplasty: Implant Design Considerations

Ujash Sheth et al. Curr Rev Musculoskelet Med. 2019 Dec.

Abstract

Purpose of review: Our understanding of the reverse total shoulder arthroplasty (RTSA) has grown exponentially since Grammont first introduced his design in 1985. There are a multitude of implant-related variables to consider when performing RTSA. The purpose of this article is to provide a review of these design considerations.

Recent findings: Current literature demonstrates that the traditional Grammont prosthesis has over 90% survivorship at 10 years. Despite these promising results, there have been concerns raised over the significant rate of scapular notching observed. As a result, the traditional RTSA design has been modified to minimize this complication and maximize impingement-free motion. Modern RTSA designs with a cementless, curved, short-stemmed eccentric onlay humeral component combined with a large, lateralized glenosphere placed in 10° of inferior tilt with > 3.5 mm of inferior overhang have been found to provide excellent results. However, all implant design features must be considered on a case-by-case basis to optimize outcome for each patient. Humeral and glenoid implant design variables have evolved as the biomechanics of RTSA have been further elucidated. Consideration of these variables allows the surgeon to maximize joint efficiency, improve impingement-free range of motion, decrease the risk of scapular notching, preserve bone stock, and minimize the risk of instability.

Keywords: Glenosphere; Humeral tray; Lateralization; Neck-shaft angle.

PubMed Disclaimer

Conflict of interest statement

Ujash Sheth declares that he has no conflict of interest.

Matthew Saltzman has received royalties from Medacta and Wright Medical.

References

    1. Administration FaD. 510(k) summary: Delta shoulder. Rockville; 2003. http://www.accessdata.fda.gov/cdrh_docs/pdf2/K021478.pdf. Accessed June 12 2019
    1. Grammont PM, Baulot E. Delta shoulder prosthesis for rotator cuff rupture. Orthopedics. 1993;16(1):65–68. doi: 10.3928/0147-7447-19930101-11. - DOI - PubMed
    1. Jain NB, Higgins LD, Losina E, Collins J, Blazar PE, Katz JN. Epidemiology of musculoskeletal upper extremity ambulatory surgery in the United States. BMC Musculoskelet Disord. 2014;15:4. doi: 10.1186/1471-2474-15-4. - DOI - PMC - PubMed
    1. Nam D, Kepler CK, Neviaser AS, Jones KJ, Wright TM, Craig EV, et al. Reverse total shoulder arthroplasty: current concepts, results, and component wear analysis. J Bone Joint Surg Am. 2010;92(Suppl 2):23–35. doi: 10.2106/JBJS.J.00769. - DOI - PubMed
    1. Gerber C, Pennington SD, Nyffeler RW. Reverse total shoulder arthroplasty. J Am Acad Orthop Surg. 2009;17(5):284–295. doi: 10.5435/00124635-200905000-00003. - DOI - PubMed