Accuracy and Reliability of Infrared Analyzers for Measuring Human Milk Macronutrients in a Milk Bank Setting
- PMID: 31723725
- PMCID: PMC6838652
- DOI: 10.1093/cdn/nzz116
Accuracy and Reliability of Infrared Analyzers for Measuring Human Milk Macronutrients in a Milk Bank Setting
Abstract
Background: Infrared (IR) analysis is an emerging technology that may be a useful tool for milk banks to manage the nutrient variability in donor human milk.
Objective: To evaluate the accuracy, reliability, and comparability of commercial infrared analyzers for measuring human milk macronutrients in a milk bank setting.
Methods: Three nonprofit milk banks received blinded test kits of human milk that had been assessed using reference methods. Four infrared instruments were used to measure macronutrients as follows: 1 filtered mid-IR, 2 Fourier-transformed full-spectra mid-IR, and 1 near-IR. Twenty-five unique samples were read concurrently for the accuracy arm. An identical sample was read daily for 1 mo for the reliability arm.
Results: Values for R 2 describing relationships with reference methods for total fat, crude protein, and lactose, were as follows: filtered mid-IR, 0.98, 0.94, and 0.48; Fourier-transformed full-spectra mid-IR, 0.97, 0.93, and 0.36 for instrument 1 and 0.98, 0.98, and 0.31 for instrument 2; and near-IR 0.93, 0.93, and 0.12. There was no significant difference between instruments for crude protein and total fat measurements. There were significant differences in carbohydrate measurements between instruments. For 1 mo of daily measurements in the reliability arm, CVs for filtered mid-IR were ≤4.6%, for Fourier-transformed full spectra mid-IR were ≤1.7%, and for near-IR were ≤5.1%.
Conclusions: Infrared analysis is an accurate and reliable method for measuring crude protein and total fat in a milk bank setting. Carbohydrate measurements are less accurate and are significantly different between instruments, which will likely lead to differences in derived calorie values.
Keywords: donor milk; human milk; infrared; macronutrients; milk bank.
Copyright © The Author(s) 2019.
Figures


Similar articles
-
Rapid measurement of macronutrients in breast milk: How reliable are infrared milk analyzers?Clin Nutr. 2015 Jun;34(3):465-76. doi: 10.1016/j.clnu.2014.05.005. Epub 2014 May 17. Clin Nutr. 2015. PMID: 24912866 Free PMC article.
-
Rapid measurement of human milk macronutrients in the neonatal intensive care unit: accuracy and precision of fourier transform mid-infrared spectroscopy.J Hum Lact. 2014 May;30(2):180-9. doi: 10.1177/0890334413517941. Epub 2014 Jan 14. J Hum Lact. 2014. PMID: 24423984
-
Infrared analysis for determining macronutrients in human milk.J Pediatr Gastroenterol Nutr. 1988 Mar-Apr;7(2):229-35. doi: 10.1097/00005176-198803000-00013. J Pediatr Gastroenterol Nutr. 1988. PMID: 3351708
-
Comparing human milk macronutrients measured using analyzers based on mid-infrared spectroscopy and ultrasound and the application of machine learning in data fitting.BMC Pregnancy Childbirth. 2022 Jul 14;22(1):562. doi: 10.1186/s12884-022-04891-w. BMC Pregnancy Childbirth. 2022. PMID: 35836199 Free PMC article.
-
Direct and indirect predictions of enteric methane daily production, yield, and intensity per unit of milk and cheese, from fatty acids and milk Fourier-transform infrared spectra.J Dairy Sci. 2018 Aug;101(8):7219-7235. doi: 10.3168/jds.2017-14289. Epub 2018 May 24. J Dairy Sci. 2018. PMID: 29803412
Cited by
-
Using Nature to Nurture: Breast Milk Analysis and Fortification to Improve Growth and Neurodevelopmental Outcomes in Preterm Infants.Nutrients. 2021 Nov 29;13(12):4307. doi: 10.3390/nu13124307. Nutrients. 2021. PMID: 34959859 Free PMC article. Review.
-
Cohort profile: Japanese human milk study, a prospective birth cohort: baseline data for lactating women, infants and human milk macronutrients.BMJ Open. 2021 Dec 30;11(12):e055028. doi: 10.1136/bmjopen-2021-055028. BMJ Open. 2021. PMID: 36282635 Free PMC article.
-
Donor human milk: the influence of processing technologies on its nutritional and microbial composition.Front Nutr. 2024 Nov 1;11:1468886. doi: 10.3389/fnut.2024.1468886. eCollection 2024. Front Nutr. 2024. PMID: 39555198 Free PMC article. Review.
-
Macronutrient concentrations in human milk beyond the first half year of lactation: a cohort study.Arch Dis Child Fetal Neonatal Ed. 2025 Apr 17;110(3):248-252. doi: 10.1136/archdischild-2024-327319. Arch Dis Child Fetal Neonatal Ed. 2025. PMID: 39532521 Free PMC article.
-
The Nutritional Composition and Energy Content of Donor Human Milk: A Systematic Review.Adv Nutr. 2020 Jul 1;11(4):960-970. doi: 10.1093/advances/nmaa014. Adv Nutr. 2020. PMID: 32119744 Free PMC article.
References
-
- Allen JC, Keller RP, Archer P, Neville MC. Studies in human lactation: milk composition and daily secretion rates of macronutrients in the first year of lactation. Am J Clin Nutr 1991;54(1):69–80. - PubMed
-
- Nommsen LA, Lovelady CA, Heinig MJ, Lönnerdal B, Dewey KG. Determinants of energy, protein, lipid, and lactose concentrations in human milk during the first 12 mo of lactation: the DARLING Study. Am J Clin Nutr 1991;53(2):457–65. - PubMed
-
- Horbar JD, Ehrenkranz RA, Badger GJ, Edwards EM, Morrow KA, Soll RF, Buzas JS, Bertino E, Gagliardi L, Bellu R. Weight growth velocity and postnatal growth failure in infants 501 to 1500 grams: 2000–2013. Pediatrics 2015;136(1):e84–92. - PubMed
-
- Committee on Nutrition, Section on Breastfeeding, Committee on Fetus and Newborn. Donor human milk for the high-risk infant: preparation, safety, and usage options in the United States. Pediatrics 2017;139(1):e20163440. - PubMed
LinkOut - more resources
Full Text Sources
Other Literature Sources