Variability of Microcirculatory Measurements in Critically Ill Patients
- PMID: 31743299
- DOI: 10.1097/SHK.0000000000001470
Variability of Microcirculatory Measurements in Critically Ill Patients
Abstract
Introduction: Monitoring the microcirculation may be helpful in guiding resuscitation in patients with circulatory shock. Sublingual side-stream dark field imaging cameras allow for noninvasive, bedside evaluation of the microcirculation, although their use in clinical practice has not yet been validated. The GlycoCheck system automatically analyzes images to determine glycocalyx thickness, red blood cell filling percentage, and vessel density. Although GlycoCheck has been used to study microcirculation in critically ill patients, little is known about the reproducibility of measurements in this population.
Materials and methods: A total of 60 critically ill patients were studied. Three consecutive microcirculation measurements were performed with the GlycoCheck system in 40 of these patients by one of two experienced observers. Twenty patients were assessed by both observers. Intra- and interobserver variability were assessed using intraclass correlation coefficients (ICCs).
Results: ICCs of single measurements were poor for glycocalyx thickness and good for filling percentage and vessel density. Reproducibility could be substantially increased for all parameters when three consecutive measurements were performed and averaged.
Discussion: GlycoCheck can be used to study microcirculation. However, to obtain reliable results three consecutive measurements should be performed and averaged. The variation of the measurements currently hampers the clinical application in individual patients.
References
-
- Levy MM, Evans LE, Rhodes A. The surviving sepsis campaign bundle: 2018 update. Crit Care Med 46 (6):997–1000, 2018.
-
- Hernandez G, Ospina-Tascon GA, Damiani LP, Estenssoro E, Dubin A, Hurtado J, Friedman G, Castro R, Alegria L, Teboul JL, et al. Effect of a resuscitation strategy targeting peripheral perfusion status vs serum lactate levels on 28-day mortality among patients with septic shock: the ANDROMEDA-SHOCK Randomized Clinical Trial. JAMA 321 (7):654–664, 2019.
-
- Goedhart PT, Khalilzada M, Bezemer R, Merza J, Ince C. Sidestream dark field (SDF) imaging: a novel stroboscopic LED ring-based imaging modality for clinical assessment of the microcirculation. Opt Express 15 (23):15101–15114, 2007.
-
- Kanoore Edul VS, Dubin A, Ince C. The microcirculation as a therapeutic target in the treatment of sepsis and shock. Semin Respir Crit Care Med 32 (5):558–568, 2011.
-
- Scorcella C, Damiani E, Domizi R, Pierantozzi S, Tondi S, Carsetti A, Ciucani S, Monaldi V, Rogani M, Marini B, et al. MicroDAIMON study: microcirculatory DAIly MONitoring in critically ill patients: a prospective observational study. Ann Intensive Care 8 (1):64, 2018.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical