A dual-center evaluation of platelet culture vials to detect the presence of microorganisms in platelets
- PMID: 31769032
- DOI: 10.1111/trf.15583
A dual-center evaluation of platelet culture vials to detect the presence of microorganisms in platelets
Abstract
Background: Microorganism contamination of platelets results in a high risk of transfusion-related sepsis. Here, the ability of culture vials (BD BACTEC Platelet Aerobic/F and Platelet Anaerobic/F vials, Becton, Dickinson and Company) to detect microorganisms in leukoreduced apheresis platelets (LRAPs) and leukoreduced whole blood platelet concentrates (LRWBPCs) was assessed.
Methods: LRAPs or LRWBPCs were inoculated into Aerobic/F and Anaerobic/F vials and placed in a blood culturing system (BD BACTEC FX System, Becton, Dickinson and Company) for growth/monitoring over 7 days to detect preexisting contamination during false-positive testing. Subsequently, platelets were seeded with microorganisms at approximately 10 CFU/mL or approximately 1 CFU/mL to simulate contamination. Aerobic/F and Anaerobic/F vials were inoculated with platelets (sets of 12). Microorganism growth was detected in the BACTEC FX instrument over 7 days. Overall, 2925 vials were tested.
Results: Of the 1905 vials included in the microorganism detection phase, 63 (3.3%) Aerobic/F and 16 (0.8%) Anaerobic/F vials were both BACTEC FX and subculture negative. From the remaining 1827 vials, two (0.1%) Anaerobic/F vials were false positive; no false positives were observed in Aerobic/F vials, and no false negatives occurred in either vial type. Of the remaining 1825 vials (99.9%), 955 Aerobic/F and 870 Anaerobic/F vials were true positives. The mean-time-to-detection range was 8.5 to 77 hours. All true-positive Aerobic/F and Anaerobic/F vials showed 100% agreement with subculture for positive identification of seeded microorganisms.
Conclusion: Aerobic/F and Anaerobic/F vials facilitate contamination detection in LRAPs and LRWBPCs down to approximately 1 CFU/mL. These results support the use of Aerobic/F and Anaerobic/F vials for quality control testing of platelets before transfusion.
© 2019 AABB.
References
REFERENCES
-
- Laki K. Our ancient heritage in blood clotting and some of its consequences. Ann N Y Acad Sci 1972;202:297-307.
-
- Kaufman RM, Djulbegovic B, Gernsheimer T, et al. Platelet transfusion: a clinical practice guideline from the AABB. Ann Intern Med 2015;162:205-13.
-
- Ellingson KD, Sapiano MRP, Haass KA, et al. Continued decline in blood collection and transfusion in the United States-2015. Transfusion 2017;57(Suppl 2):1588-98.
-
- Food and Drug Administration. Bacterial risk control strategies for blood collection establishments and transfusion services to enhance the safety and availability of platelets for transfusion. Draft guidance for industry. 2018. Available from: https://www.fda.gov/media/123448/download.
-
- Brecher ME, Hay SN. Bacterial contamination of blood components. Clin Microbiol Rev 2005;18:195-204.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical