Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2019 Nov 28;7(1):coz079.
doi: 10.1093/conphys/coz079. eCollection 2019.

Point-of-care testing for measuring haemolymph glucose in invertebrates is not a valid method

Affiliations

Point-of-care testing for measuring haemolymph glucose in invertebrates is not a valid method

Silas C Principe et al. Conserv Physiol. .

Abstract

Blood glucose is widely used as a physiological parameter for vertebrates and invertebrates. However, its measurement in the field is often difficult due to the need for expensive and non-portable equipment. Point-of-care (POC) devices, originally intended for human use, are increasingly being used for measuring blood parameters of animals in the field. In this regard, POC glucose meters are becoming valuable tools for conservation physiologists, as glucose can be a useful indicator of stress response. In invertebrates, the use of POC glucose meters is still scarce, and no study yet has evaluated their usability in crustaceans and molluscs. We tested if a POC device can be used to measure haemolymph glucose in two widely used models, Leptuca thayeri and Perna perna, compared with a standard laboratory method. The device was unable to measure glucose in P. perna haemolymph due to equipment inaccuracy and low glucose concentration in this species (10.13 ± 6.25 mg/dL). Additionally, despite the device being capable of measuring glucose in L. thayeri haemolymph, Bland-Altman plots showed a strong bias and wide limits of agreement, and Lin's concordance correlation coefficient showed a weak concordance between methods. When simulating experimental conditions, POC results differed from those found using the standard method. We conclude that POC glucose meters are unsuitable for assessing glucose in mussels and should not be used in crabs as results are inaccurate.

Keywords: Leptuca thayeri; Perna perna; fiddler crab; glucose; molluscs; portable meter.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Comparison between methods of glucose measurement (a) Linear regression of fiddler crab haemolymph glucose measured using a POC device and via the laboratory method. The grey dotted line is the expected line of perfect agreement, and the solid line is the estimated slope. (b) Bland–Altman plot of glucose measurements via a POC device and the laboratory method. Solid grey line represents bias, and dotted lines represent limits of agreement (95%).
Figure 2
Figure 2
Glucose levels of L. thayeri exposed to control and high temperature conditions. Samples were read using two different methods (laboratory method (LAB) and POC device), and comparisons between treatments for each method are shown in the above Cumming estimation plot. Glucose values are plotted on the upper axes with vertical lines indicating mean (as a gap) ± SD. Each mean difference is plotted on the lower axes as a bootstrap sampling distribution. Mean differences are depicted as dots; 95% CIs are indicated by the ends of the vertical error bars. Significant difference was found just for glucose levels obtained via the laboratory method (P < 0.05).

References

    1. Aliko V, Hajdaraj G, Caci A, Faggio C (2015) Copper-induced lysosomal membrane destabilisation in haemolymph cells of mediterranean green crab (Carcinus aestuarii, Nardo, 1847) from the Narta Lagoon (Albania). Braz Arch Biol Technol 58: 750–756.
    1. Allender MC, Schumacher J, George R, Milam J, Agricola O (2010) The effects of short- and long-term hypoxia on hemolymph gas values in the American horseshoe crab (Limulus polyphemus) using a point-of-care analyzer. J Zoo Wildl Med 41: 193–200. - PubMed
    1. Altman DG, Bland JM (1983) Measurement in medicine: the analysis of method comparison studies. Statistician 32: 307–317.
    1. Amachree D, Moody AJ, Handy RD (2014) Comparison of intermittent and continuous exposures to inorganic mercury in the mussel, Mytilus edulis: accumulation and sub-lethal physiological effects. Ecotoxicol Environ Saf 109: 133–142. - PubMed
    1. Andrewartha SJ, Munns SL, Edwards A (2016) Calibration of the HemoCue point-of-care analyser for determining haemoglobin concentration in a lizard and a fish. Conserv Physiol 4: 1–6. - PMC - PubMed