Abstract
Introduction: Powered robotic exoskeletons are a promising solution to enable standing and walking in patients with spinal cord injury (SCI). Although training and walking with an exoskeleton in motor complete SCI patients is considered safe, the risks of unexpected (technical) adverse events and the risk of fractures are not fully understood. This article reports the occurrence of two different cases of bone fracture during exoskeleton usage. Furthermore, advice is given for extra safety training and instructions.
Case presentation: The first case concerns a 47-year-old woman with T12 AIS A SCI. Her exoskeleton shut down unexpectedly probably causing a misalignment of the joints of her lower extremities relative to the joints of the exoskeleton, which resulted in a fracture of her left tibia. The second case involves a 39-year-old man with L1 AIS B SCI. An unexpected fracture of the right distal tibia occurred without a specific prior (traumatic) incident.
Discussion: Exoskeleton training instructors, SCI patients and their buddies should be instructed how to handle emergency situations. Furthermore, they should be aware of the risk of stress fractures of the lower extremities. Proper alignment of the exoskeleton relative to the body is of utmost importance to reduce fracture risk. In the case of swelling and discoloring of the skin, radiographic examination should be performed in order to exclude any fracture.
Keywords: Neurological disorders; Risk factors; Skeleton; Therapeutics.
© International Spinal Cord Society 2019.
Conflict of interest statement
Conflict of interestThe authors declare that they have no conflict of interest.
Figures
Similar articles
-
Bilateral tibial fractures associated with powered exoskeleton use in complete spinal cord injury - a case report & literature review.Spinal Cord Ser Cases. 2024 Apr 16;10(1):22. doi: 10.1038/s41394-024-00635-4. Spinal Cord Ser Cases. 2024. PMID: 38627367 Free PMC article. Review.
-
Locomotor training using an overground robotic exoskeleton in long-term manual wheelchair users with a chronic spinal cord injury living in the community: Lessons learned from a feasibility study in terms of recruitment, attendance, learnability, performance and safety.J Neuroeng Rehabil. 2018 Mar 1;15(1):12. doi: 10.1186/s12984-018-0354-2. J Neuroeng Rehabil. 2018. PMID: 29490678 Free PMC article. Clinical Trial.
-
Assistive powered exoskeleton for complete spinal cord injury: correlations between walking ability and exoskeleton control.Eur J Phys Rehabil Med. 2019 Apr;55(2):209-216. doi: 10.23736/S1973-9087.18.05308-X. Epub 2018 Aug 27. Eur J Phys Rehabil Med. 2019. PMID: 30156088
-
Cardiopulmonary function after robotic exoskeleton-assisted over-ground walking training of a patient with an incomplete spinal cord injury: Case report.Medicine (Baltimore). 2019 Dec;98(50):e18286. doi: 10.1097/MD.0000000000018286. Medicine (Baltimore). 2019. PMID: 31852105 Free PMC article.
-
A systematic review of the determinants of implementation of a locomotor training program using a powered exoskeleton for individuals with a spinal cord injury.Clin Rehabil. 2023 Aug;37(8):1119-1138. doi: 10.1177/02692155231164092. Epub 2023 Apr 10. Clin Rehabil. 2023. PMID: 37036438 Free PMC article.
Cited by
-
Exoskeleton home and community use in people with complete spinal cord injury.Sci Rep. 2020 Sep 24;10(1):15600. doi: 10.1038/s41598-020-72397-6. Sci Rep. 2020. PMID: 32973244 Free PMC article.
-
Bilateral tibial fractures associated with powered exoskeleton use in complete spinal cord injury - a case report & literature review.Spinal Cord Ser Cases. 2024 Apr 16;10(1):22. doi: 10.1038/s41394-024-00635-4. Spinal Cord Ser Cases. 2024. PMID: 38627367 Free PMC article. Review.
-
Effects of an active ankle exoskeleton on the walking biomechanics of healthy men.Front Bioeng Biotechnol. 2025 Mar 14;13:1533001. doi: 10.3389/fbioe.2025.1533001. eCollection 2025. Front Bioeng Biotechnol. 2025. PMID: 40161516 Free PMC article.
-
Safety Assessment of Rehabilitation Robots: A Review Identifying Safety Skills and Current Knowledge Gaps.Front Robot AI. 2021 Mar 22;8:602878. doi: 10.3389/frobt.2021.602878. eCollection 2021. Front Robot AI. 2021. PMID: 33937345 Free PMC article. Review.
-
Physiotherapy using a free-standing robotic exoskeleton for patients with spinal cord injury: a feasibility study.J Neuroeng Rehabil. 2021 Dec 25;18(1):180. doi: 10.1186/s12984-021-00967-4. J Neuroeng Rehabil. 2021. PMID: 34953501 Free PMC article.
References
-
- Gagnon DH, Escalona MJ, Vermette M, Carvalho LP, Karelis AD, Duclos C, et al. Locomotor training using an overground robotic exoskeleton in long-term manual wheelchair users with a chronic spinal cord injury living in the community: Lessons learned from a feasibility study in terms of recruitment, attendance, learnability, performa. J Neuroeng Rehabil. 2018;15:12. doi: 10.1186/s12984-018-0354-2. - DOI - PMC - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical