Evaluation of OptiFlow™-MS/MS for bioanalysis of pharmaceutical drugs and metabolites
- PMID: 31849261
- DOI: 10.4155/bio-2019-0250
Evaluation of OptiFlow™-MS/MS for bioanalysis of pharmaceutical drugs and metabolites
Abstract
Aim: Microflow tandem mass spectrometry-based methods have been proposed as options to improve sensitivity and selectivity while improving sample utility and solvent consumption. Here, we evaluate a newly introduced microflow source, OptiFlow™, for quantitative performance. Results/methodology: We performed a comparison of the OptiFlow and IonDrive™ sources, respectively, on the same triple quadrupole mass spectrometer. The comparison used a neat cocktail of commercially available drugs and extracted plasma samples monitoring midazolam and alprazolam metabolites. Microflow produced a 2-4× signal increase for the neat drug cocktail and a 5-10× increase for extracted plasma samples. Conclusion: The OptiFlow method consistently gave increased signal response relative to the IonDrive method and enabled a better lower limit of quantitation for defining phamacokinetics.
Keywords: LC–MS/MS; microflow; microsampling; pharmacokinetics.