Percutaneous closure of restrictive-type perimembranous ventricular septal defect using the new KONAR multifunctional occluder: Midterm outcomes of the first middle-eastern experience
- PMID: 31886940
- DOI: 10.1002/ccd.28678
Percutaneous closure of restrictive-type perimembranous ventricular septal defect using the new KONAR multifunctional occluder: Midterm outcomes of the first middle-eastern experience
Abstract
Objectives: To evaluate the safety, efficiency, and midterm outcomes of transcatheter perimembranous ventricular septal defect (pmVSD) closure using the new KONAR-MF™ VSD occluder.
Background: Off-label percutaneous pmVSD closure is a well-established procedure with promising results. Yet, interventionists are still searching for the ideal device.
Methods: Between June 2018 and March 2019, 20 patients with hemodynamically significant but restrictive-type pmVSD underwent an attempted transcatheter closure using the new KONAR-MF™. All implantations were performed retrogradely under general anesthesia, transoesophageal echocardiography, and fluoroscopic guidance. Prospective follow-up using transthoracic echocardiography and electrocardiogram was done until August 2019.
Results: The median age was 6.4 years (8 months to 43.4 years), and the median body weight was 17.3 (9-74) kg. The mean defect size on the left ventricular aspect was 11.7 ± 2.8 mm. All devices were successfully and rapidly implanted. One device embolized in the pulmonary artery, 24 hr after implantation and was percutaneously retrieved with no complication. Over a mean follow-up period of 8.2 ± 3.0 months, complete closure was achieved in 84.2% of patients. One new onset of mild aortic regurgitation was detected before discharge and remained stable. Tricuspid valve regurgitation, complete heart block, major complication, and death were not observed.
Conclusions: Transcatheter pmVSD closure using the KONAR-MF™ can be successfully performed in adult and pediatric patients. It is a safe and promising device, designed to provide high conformability to septal defects with a lower risk of heart block. Retrograde implantation allows procedural flexibility, efficient deliverability, and control of valvular interferences.
Keywords: Lifetech; congenital heart defect; device closure; percutaneous.
© 2019 Wiley Periodicals, Inc.
References
REFERENCES
-
- Morray BH. Ventricular septal defect closure devices, techniques, and outcomes. Interv Cardiol Clin. 2019;8(1):1-10.
-
- Hong ZN, Chen Q, Huang LQ, Cao H. A meta-analysis of perventricular device closure of perimembranous ventricular septal defect. J Cardiothorac Surg. 2019;14(1):119.
-
- Santhanam H, Yang L, Chen Z, Tai BC, Rajgor DD, Quek SC. A meta-analysis of transcatheter device closure of perimembranous ventricular septal defect. Int J Cardiol. 2018;254:75-83.
-
- Yang L, Tai BC, Khin LW, Quek SC. A systematic review on the efficacy and safety of transcatheter device closure of ventricular septal defects (VSD). J Interv Cardiol. 2014;27(3):260-272.
-
- Haddad RN, Daou L, Saliba Z. Device closure of perimembranous ventricular septal defect: choosing between amplatzer occluders. Front Pediatr. 2019;7:300.
MeSH terms
LinkOut - more resources
Full Text Sources