Assessment of Adult Diagnostic Reference Levels for Panoramic Radiography in Tamil Nadu Region
- PMID: 31908390
- PMCID: PMC6936199
- DOI: 10.4103/jmp.JMP_77_19
Assessment of Adult Diagnostic Reference Levels for Panoramic Radiography in Tamil Nadu Region
Abstract
Aim: The aim of this study was to calculate dose area product (DAP) and to determine diagnostic reference level (DRL) for adult panoramic procedures in Tamil Nadu.
Materials and methods: In this study, air kerma on the front side of the secondary collimator was measured with a Black Piranha, RTI Electronics, Sweden and multiplied with the corresponding exposed area to calculate DAP. The obtained DAP values were further analyzed, and DRL was calculated using the Microsoft Excel software. The study was carried out with regular adult exposure parameters.
Results: The mean, range, and 3rd quartile values for 67 panoramic scanners in Tamil Nadu, India, were calculated as 94 mGycm2, 41 mGycm2-165 mGycm2, and 114.3 mGycm2, respectively. The results are comparable with other international studies.
Conclusion: The present study suggests that further optimization can be achieved in many centers by the recruitment of professionally qualified radiographers and conducting periodic training programs on the optimization of exposure parameters. Considering this as the first study for the dental DRL assessment, further studies are suggested to establish national dental DRL in India.
Keywords: Dental radiography; diagnostic reference level; dose area product; radiation protection.
Copyright: © 2019 Journal of Medical Physics.
Conflict of interest statement
There are no conflicts of interest.
Figures







References
-
- Kaeppler G, Dietz K, Herz K, Reinert S. Factors influencing the absorbed dose in intraoral radiography. Dentomaxillofac Radiol. 2007;36:506–13. - PubMed
-
- Hujoel P, Hollender L, Bollen AM, Young JD, McGee M, Grosso A. Head-and-neck organ doses from an episode of orthodontic care. Am J Orthod Dentofacial Orthop. 2008;133:210–7. - PubMed
-
- Vañó E, Miller DL, Martin CJ, Rehani MM, Kang K, Rosenstein M, et al. ICRP publication 135: Diagnostic reference levels in medical imaging. Ann ICRP. 2017;46:1–44. - PubMed
-
- International Atomic Energy Agency. Diagnostic Reference Levels (DRLs) in Medical Imaging. 2018. [Last accessed on 2018 Jun 15]. Available from: https://www.iaea.org/resources/rpop/health-professionals/nuclear-medicin... .
-
- Lee JS, Kim YH, Yoon SJ, Kang BC. Reference dose levels for dental panoramic radiography in Gwangju, South Korea. Radiat Prot Dosimetry. 2010;142:184–90. - PubMed