In vitro comparison of aortic heart valve prostheses. Part 1: Mechanical valves
- PMID: 3193804
In vitro comparison of aortic heart valve prostheses. Part 1: Mechanical valves
Abstract
The hydrodynamic performance of a series of mechanical heart valve prostheses is measured in the aortic position. All experiments are performed in an electrohydraulic, computer-controlled pulse duplicator simulating the left side of the human circulatory system. Testing conditions are set according to a Food and Drug Administration interlaboratory comparison protocol with cardiac outputs of 3.0, 4.5, 6.5, and 8.0 L/min at a constant heart rate of 70 beats/min. Mean systolic pressure differences, volume losses, and energy losses, as well as dimensionless pressure losses and energy loss coefficients, are calculated from the recorded pressure, volume, and flow tracings. The results of 10 mechanical heart valve prostheses (eight tilting disc valves and two bileaflet valves) are presented and their clinical relevance is discussed.
Similar articles
-
In vitro comparison of bileaflet aortic heart valve prostheses. St. Jude Medical, CarboMedics, modified Edwards-Duromedics, and Sorin-Bicarbon valves.J Thorac Cardiovasc Surg. 1993 Sep;106(3):412-20. J Thorac Cardiovasc Surg. 1993. PMID: 8361181
-
In vitro flow dynamics of a new mechanical cardiac valve prosthesis--"tricusp".Scand J Thorac Cardiovasc Surg. 1996;30(3-4):117-23. doi: 10.3109/14017439609107256. Scand J Thorac Cardiovasc Surg. 1996. PMID: 8976031
-
Small aortic annulus: the hydrodynamic performances of 5 commercially available bileaflet mechanical valves.J Thorac Cardiovasc Surg. 2004 Sep;128(3):457-62. doi: 10.1016/j.jtcvs.2004.03.021. J Thorac Cardiovasc Surg. 2004. PMID: 15354108
-
New flexible polymeric heart valve prostheses for the mitral and aortic positions.Heart Surg Forum. 2004;7(5):E525-32. doi: 10.1532/HSF98.20041083. Heart Surg Forum. 2004. PMID: 15799940 Review.
-
Fluid-structure interaction simulation of mechanical aortic valves: a narrative review exploring its role in total product life cycle.Front Med Technol. 2024 Jul 1;6:1399729. doi: 10.3389/fmedt.2024.1399729. eCollection 2024. Front Med Technol. 2024. PMID: 39011523 Free PMC article. Review.
Cited by
-
Computer-controlled in vitro model of the human left heart.Med Biol Eng Comput. 1992 Nov;30(6):656-9. doi: 10.1007/BF02446800. Med Biol Eng Comput. 1992. PMID: 1297023
-
Comparison of Aspiration Catheters with Modified Standard Catheters for Treatment of Large Pulmonary Embolism Using an In-vitro Patho-Physiological Model.Cardiovasc Intervent Radiol. 2022 Jan;45(1):112-120. doi: 10.1007/s00270-021-02987-y. Epub 2021 Nov 18. Cardiovasc Intervent Radiol. 2022. PMID: 34796375 Free PMC article.