Evaluation of image quality and radiation dose in low tube voltage coronary computed tomography angiography
- PMID: 31949446
- PMCID: PMC6954360
- DOI: 10.22122/arya.v15i5.1406
Evaluation of image quality and radiation dose in low tube voltage coronary computed tomography angiography
Abstract
Background: Coronary computed tomography angiography (CCTA) is an important modality in diagnosis of coronary artery disease (CAD). Owing to the fact that computed tomography (CT) examinations are performed using ionizing radiation; applying radiation dose-reduction strategies seems to be necessary. Lowering tube voltage (in kV) according to the patient's body mass index (BMI) or weight is an approach that is investigated by many researchers. The goal of this study was to evaluate the impact of low tube voltage CCTA on radiation dose and image quality in order to decrease radiation dose in selected patients who meet inclusion criteria of the introduced protocol.
Methods: Patients with clinical indications of CCTA who met inclusion criteria were classified in two groups randomly. Imaging of two groups was performed using 120 kV and 100 kV, respectively. Subjective and objective parameters of image quality and radiation dose of two groups were measured. Afterward, data were analyzed by appropriate statistical tests using SPSS software.
Results: While differences in image quality between two groups were not significant, radiation dose of patients who underwent 100 kV CCTA was significantly lower than the other group. Effective doses (EDs) of first and second groups were 22.30 ± 5.48 mSv and 13.82 ± 2.00 mSv, respectively (P < 0.001).
Conclusion: Lowering tube voltage in non-obese patients is an effective and practical approach to radiation dose reduction without missing image quality that should be considered especially for female patients.
Keywords: Computed Tomography Angiography; Contrast Agent; Coronary Angiography; Image Enhancement; Radiation Dosage; X-Rays.
© 2019 Isfahan Cardiovascular Research Center & Isfahan University of Medical Sciences.
Figures
References
-
- Shah A, Das P, Subkovas E, Buch AN, Rees M, Bellamy C. Radiation dose during coronary angiogram: Relation to body mass index. Heart Lung Circ. 2015;24(1):21–5. - PubMed
-
- Kayan M, Koroglu M, Yesildag A, Ceylan E, Aktas AR, Yasar S, et al. Carotid CT-angiography: Low versus standard volume contrast media and low kV protocol for 128-slice MDCT. Eur J Radiol. 2012;81(9):2144–7. - PubMed
-
- Shapiro BP, Young PM, Kantor B, Choe YH, McCollough CH, Gerber TC. Radiation dose reduction in CT coronary angiography. Curr Cardiol Rep. 2010;12(1):59–67. - PubMed
-
- Xu L, Zhang Z. Coronary CT angiography with low radiation dose. Int J Cardiovasc Imaging. 2010;26(Suppl 1):17–25. - PubMed
LinkOut - more resources
Full Text Sources
Miscellaneous