Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2020 Sep;29(9):717-726.
doi: 10.1136/bmjqs-2019-010368. Epub 2020 Jan 20.

Time series evaluation of improvement interventions to reduce alarm notifications in a paediatric hospital

Affiliations

Time series evaluation of improvement interventions to reduce alarm notifications in a paediatric hospital

Colleen M Pater et al. BMJ Qual Saf. 2020 Sep.

Abstract

BACKGROUND10: The Joint Commission identified inpatient alarm reduction as an opportunity to improve patient safety; enhance patient, family and nursing satisfaction; and optimise workflow. We used quality improvement (QI) methods to safely decrease non-actionable alarm notifications to bedside providers.

Methods: In a paediatric tertiary care centre, we convened a multidisciplinary team to address alarm notifications in our acute care cardiology unit. Alarm notification was defined as any alert to bedside providers for each patient-triggered monitor alarm. Our aim was to decrease alarm notifications per monitored bed per day by 60%. Plan-Do-Study-Act testing cycles included updating notification technology, establishing alarm logic and modifying bedside workflow processes, including silencing the volume on all bedside monitors. Our secondary outcome measure was nursing satisfaction. Balancing safety measures included floor to intensive care unit transfers and patient acuity level.

Results: At baseline, there was an average of 71 initial alarm notifications per monitored bed per day. Over a 3.5-year improvement period (2014-2017), the rate decreased by 68% to 22 initial alarm notifications per monitored bed per day. The proportion of initial to total alarm notifications remained stable, decreasing slightly from 51% to 40%. There was a significant improvement in subjective nursing satisfaction. At baseline, 32% of nurses agreed they were able to respond to alarms appropriately and quickly. Following interventions, agreement increased to 76% (p<0.001). We sustained these improvements over a year without a change in monitored balancing measures.

Conclusion: We successfully reduced alarm notifications while preserving patient safety over a 4-year period in a complex paediatric patient population using technological advances and QI methodology. Continued efforts are needed to further optimise monitor use across paediatric hospital units.

Keywords: alarm reduction; quality improvement.