Very high pacing thresholds during long-term follow-up predicted by a combination of implant pacing threshold and impedance in leadless transcatheter pacemakers
- PMID: 31967367
- DOI: 10.1111/jce.14360
Very high pacing thresholds during long-term follow-up predicted by a combination of implant pacing threshold and impedance in leadless transcatheter pacemakers
Abstract
Background: Micra transcatheter pacemaker system (TPS) usually achieves low implant pacing threshold (IPT). However, IPT may increase in some patients during follow-up.
Aim: To apply implant parameters in predicting long-term occurrence of very high pacing threshold (VHPT) in patients with Micra-TPS.
Methods: A cohort of 110 consecutive patients implanted with a Micra-TPS from 2014 to 2018 was evaluated at discharge and at 1, 12, 24, 36, and 48 months follow-up. VHPT was defined as greater than 2 V/0.24 ms. VHPT predictors were identified.
Results: Micra-TPS was implanted successfully in 108 patients (98.2%). During a mean follow-up of 24 ± 16 months, 18 patients (16.7%) died of causes nonpacemaker-related, and 4 (3.8%) developed VHPT. Patients with VHPT had higher IPT and lower implant impedance than patients with non-VHPT: 1 ± 0.31 vs 0.55 ± 0.29 V/0.24 ms (P = .003) and 580 ± 59 vs 837 ± 232 Ω (P = .03), respectively. IPT and impedance had excellent discriminative power to predict VHPT (area under the curve: 0.85 ± 0.07 and 0.91 ± 0.05, respectively). Negative predictive value (NPV) of IPT ≤ 0.5 V/0.24 ms was 100%; positive predictive value (PPV) was 8% throughout follow-up. Implant impedance ≤ 600 Ω had NPV of 99% throughout follow-up, whereas PPV varied: 16%, 21%, 16%, and 28% at 1, 12, 24, and 36 months, respectively. Sequential combination of IPT greater than 0.5 V/0.24 ms and impedance ≤ 600 Ω improved PPV to 25%, 35%, 27%, and 44%, respectively, whereas NPV remained 99% throughout follow-up.
Conclusion: Despite favorable long-term electrical performance of Micra-TPS, a small percent of patients developed VHPT during follow-up. A sequential combination of IPT and impedance could allow the implanter to identify patients who will develop VHPT during long-term follow-up.
Keywords: Leadless transcatheter pacemaker; Pacing thresholds; battery drain; impedance.
© 2020 Wiley Periodicals, Inc.
References
REFERENCES
-
- Ritter P, Duray GZ, Zhang S, et al. The rationale and design of the Micra transcatheter pacing study: safety and efficacy of a novel miniaturized pacemaker. Europace. 2015;17:807-813.
-
- Early performance of a miniaturized leadless cardiac pacemaker: the Micra transcatheter pacing study. Eur Heart J. 2015;36:2510-2519.
-
- Reynolds D, Duray GZ, Omar R, et al. A leadless intracardiac transcatheter pacing system. N Engl J Med. 2016;374:533-541.
-
- Duray GZ, Ritter P, El-Chami M, et al. Long-term performance of a transcatheter pacing system: 12-month results from the Micra transcatheter pacing study. Heart Rhythm. 2017;14:702-709.
-
- Aytemir K, Canpolat U, Yorgun H, et al. Usefulness of ‘figure-of-eight' suture to achieve haemostasis after removal of 15-French calibre femoral venous sheath in patients undergoing cryoablation. Europace. 2016;18:1545-1550.