Multicenter Study
. 2020 Aug;47(9):2175-2185.
doi: 10.1007/s00259-020-04689-y. Epub 2020 Jan 25.

A kinetics-based approach to amyloid PET semi-quantification

Affiliations
Multicenter Study

A kinetics-based approach to amyloid PET semi-quantification

A Chincarini et al. Eur J Nucl Med Mol Imaging. 2020 Aug.

Abstract

Purpose: To develop and validate a semi-quantification method (time-delayed ratio, TDr) applied to amyloid PET scans, based on tracer kinetics information.

Methods: The TDr method requires two static scans per subject: one early (~ 0-10 min after the injection) and one late (typically 50-70 min or 90-100 min after the injection, depending on the tracer). High perfusion regions are delineated on the early scan and applied onto the late scan. A SUVr-like ratio is calculated between the average intensities in the high perfusion regions and the late scan hotspot. TDr was applied to a naturalistic multicenter dataset of 143 subjects acquired with [18F]florbetapir. TDr values are compared to visual evaluation, cortical-cerebellar SUVr, and to the geometrical semi-quantification method ELBA. All three methods are gauged versus the heterogeneity of the dataset.

Results: TDr shows excellent agreement with respect to the binary visual assessment (AUC = 0.99) and significantly correlates with both validated semi-quantification methods, reaching a Pearson correlation coefficient of 0.86 with respect to ELBA.

Conclusions: TDr is an alternative approach to previously validated ones (SUVr and ELBA). It requires minimal image processing; it is independent on predefined regions of interest and does not require MR registration. Besides, it takes advantage on the availability of early scans which are becoming common practice while imposing a negligible added patient discomfort.

Keywords: Amyloid PET; Fluorinated tracer; Semi-quantification; Tracer kinetics.