Orbscan 3 Versus Pentacam HR: Evaluating the Possible Interchangeable Use of Various Parameters

Affiliations

Orbscan 3 Versus Pentacam HR: Evaluating the Possible Interchangeable Use of Various Parameters

Hesham Mohamed Gharieb et al. Cornea. 2020 May.

Abstract

Purpose: Many topography/tomography devices are available, and their interchangeable use is frequently questioned. This study aimed to compare the analogous indices of Orbscan 3 and Pentacam HR to detect whether the interchangeable use of the 2 devices is feasible.

Methods: This retrospective study was conducted at Eye World Hospital, Giza, Egypt. Displays of Orbscan 3 (Bausch & Lomb; Technolas Perfect Vision GmbH) and Pentacam HR (Oculus Optikegrate) for 660 healthy right eyes were explored for various analogous indices. Bland-Altman plots with 95% limits of agreement (LoA) were used for comparisons.

Results: The highest systematic error and lowest agreement existed in the front and back elevations from the best fit sphere (mean differences of 4.7 and 9.2 μm, respectively, and 95% LoA of -2.5 to 12.0 and -10.0 to 28.6, respectively). There was a relatively wide 95% LoA for each pachymetry at the corneal apex (-18 to 22 μm) and at the thinnest location (-22 to 22 μm), anterior chamber depth (-0.88 to 0.95 mm), flattest keratometry (-0.44 to 0.62 D), and steepest keratometry (-0.71 to 0.89 D). By contrast, the lowest systematic error and highest agreement existed for front and back best fit sphere radii of curvature (mean difference of 0.08 mm and 95% LoA of -0.08 to 0.24 and mean difference of 0.03 mm and 95% LoA of -0.17 to 0.23, respectively).

Conclusions: This study negates the accuracy of the use of devices interchangeably regarding most parameters. Therefore, examinations with the 2 devices should not be regarded comparable.

PubMed Disclaimer

References

    1. Crawford AZ, Patel DV, McGhee CN. Comparison and repeatability of keratometric and corneal power measurements obtained by Orbscan II, Pentacam, and Galilei corneal tomography systems. Am J Ophthalmol. 2013;156:53–60.
    1. Sanhermelando MV, Lleó A, Alonso L, et al. Repeatability of central corneal thickness and ocular anterior chamber depth measurements with the orbscan topography system. Eur J Anat. 2002;6:59–64.
    1. Wei RH, Lim L, Chan WK, et al. Evaluation of Orbscan II corneal topography in individuals with myopia. Ophthalmology. 2006;113:177–183.
    1. Orbscan IIz, Anterior segment analysis system: portaloptico. 2002. Available at: https://www.portaloptico.com.mx/descargas/technolas-perfect-vision/180-c.... Accessed January 3, 2019.
    1. Zyoptix ® Diagnostic Workstation 3: Promedwork 2017 [cited 2017]. Available at: https://promedwork.com/site/pt/zyoptix-diagnostic-workstation3/. Accessed January 3, 2019.