Double-tuned 31 P/1 H human head array with high performance at both frequencies for spectroscopic imaging at 9.4T
- PMID: 32003031
- DOI: 10.1002/mrm.28176
Double-tuned 31 P/1 H human head array with high performance at both frequencies for spectroscopic imaging at 9.4T
Abstract
Purpose: To develop a robust design of a human head double-tuned 31 P/1 H array, which provides good performance at both 31 P and 1 H frequencies for MR spectroscopic imaging at 9.4T.
Methods: Increasing the number of surface loops in a human head array improves the peripheral signal-to-noise ratio (SNR), while the central SNR doesn't substantially change. High peripheral SNR can contaminate MR spectroscopic imaging data at both 1 H and 31 P frequency. To minimize this effect, we limited the number of elements in the 31 P array to 10, i.e., 8 transceiver surface loops circumscribing the head and 2 receive "vertical" loops placed at the superior location. The 1 H-portion of the array also consists of 10 elements, i.e., 8 transceiver surface loops circumscribing the head and 2 transceiver "vertical" loops at the superior location of the head. Both the 31 P array and 1 H array are placed in a single layer at the same distance to the head, which provides high loading and, thus, a good performance for both arrays.
Results: Transmit efficiency of the 1 H-portion of the double-tuned array was very similar to that of the single-tuned arrays of similar size. Also, addition of the cross-loops substantially improved the brain coverage.
Conclusion: We developed a novel 31 P/1 H double-tuned array for MR spectroscopic imaging of a human brain at 9.4T. Placing both 31 P and 1 H loops in a single layer provides for high transmit efficiency at both frequencies without compromising SNR near the brain center at the 31 P-frequency. Addition of the cross-loops at the superior location improves the brain coverage.
Keywords: MRSI; RF human head coil; central SNR; double-tuned arrays; ultra-high field MRI.
© 2020 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine.
References
REFERENCES
-
- Öz G, Alger JR, Barker PB, et al. Clinical proton MR spectroscopy in central nervous system disorders. Radiology. 2014;270:658-679.
-
- Schnall MD, Subramanian VH, Leigh JS, Chance B. A new double-tuned probed for concurrent 1H and 31P NMR. J Magn Reson. 1985;65:122.
-
- Avdievich NI. Transceiver phased arrays for human brain studies at 7 T. Appl Magn Reson. 2011;41:483-506.
-
- Barberi EA, Gati JS, Rutt BK, Menon RS. A transmit-only/receive-only (ToRo) RF system for high-field MRI/MRS applications. Magn Reson Med. 2000;43:284-289.
-
- Vaughan JT, Adriany G, Garwood M, et al. Detunable Transverse Electromagnetic (TEM) volume coil for high-field NMR. Magn Reson Med. 2002;47:990-1000.
Publication types
MeSH terms
LinkOut - more resources
Other Literature Sources
Medical