. 2020 Feb 25;58(3):375-383.
doi: 10.1515/cclm-2019-0534.

Highly accurate and explainable detection of specimen mix-up using a machine learning model

Affiliations

Highly accurate and explainable detection of specimen mix-up using a machine learning model

Tomohiro Mitani et al. Clin Chem Lab Med. .

Abstract

Background Delta check is widely used for detecting specimen mix-ups. Owing to the inadequate specificity and sparseness of the absolute incidence of mix-ups, the positive predictive value (PPV) of delta check is considerably low as it is labor consuming to identify true mix-up errors among a large number of false alerts. To overcome this problem, we developed a new accurate detection model through machine learning. Methods Inspired by delta check, we decided to conduct comparisons with the past examinations and broaden the time range. Fifteen common items were selected from complete blood cell counts and biochemical tests. We considered examinations in which ≥11 among the 15 items were measured simultaneously in our hospital; we created individual partial time-series data of the consecutive examinations with a sliding window size of 4. The last examinations of the partial time-series data were shuffled to generate artificial mix-up cases. After splitting the dataset into development and validation sets, we allowed a gradient-boosting-decision-tree (GBDT) model to learn using the development set to detect whether the last examination results of the partial time-series data were artificial mixed-up results. The model's performance was evaluated on the validation set. Results The area under the receiver operating characteristic curve (ROC AUC) of our model was 0.9983 (bootstrap confidence interval [bsCI]: 0.9983-0.9985). Conclusions The GBDT model was more effective in detecting specimen mix-up. The improved accuracy will enable more facilities to perform more efficient and centralized mix-up detection, leading to improved patient safety.

Keywords: anomaly detection; data-mining; delta-check method; laboratory information system; machine learning; patient safety.

PubMed Disclaimer

References

    1. Lippi G, Chance JJ, Church S, Dazzi P, Fontana R, Giavarina D, et al. Preanalytical quality improvement: from dream to reality. Clin Chem Lab Med 2011;49:1113–26.
    1. Lippi G, Blanckaert N, Bonini P, Green S, Kitchen S, Palicka V, et al. Causes, consequences, detection, and prevention of identification errors in laboratory diagnostics. Clin Chem Lab Med 2009;47:143–53.
    1. Simundic A-M, Church S, Cornes MP, Grankvist K, Lippi G, Nybo M, et al. Compliance of blood sampling procedures with the CLSI H3-A6 guidelines: an observational study by the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) working group for the preanalytical phase (WG-PRE). Clin Chem Lab Med 2015;53:1321–31.
    1. Yamashita T, Ichihara K, Miyamoto A. A novel weighted cumulative delta-check method for highly sensitive detection of specimen mix-up in the clinical laboratory. Clin Chem Lab Med 2013;51:781–9.
    1. Randell EW, Yenice S. Delta checks in the clinical laboratory. Crit Rev Clin Lab Sci 2019;11:1–23.

LinkOut - more resources