Intervendor variability of carotid intima-media thickness measurement: validation study using newly developed ultrasound phantom
- PMID: 32067178
- DOI: 10.1007/s10396-019-00995-7
Intervendor variability of carotid intima-media thickness measurement: validation study using newly developed ultrasound phantom
Erratum in
-
Correction to: Intervendor variability of carotid intima-media thickness measurement: validation study using newly developed ultrasound phantom.J Med Ultrason (2001). 2020 Jul;47(3):487-488. doi: 10.1007/s10396-020-01037-3. J Med Ultrason (2001). 2020. PMID: 32607641
Abstract
Purpose: Ultrasonography-derived carotid artery intima-media thickness (IMT) has been established as an early atherosclerotic imaging biomarker. The IMT reference value of a healthy person is approximately 0.1 × (every 10 years of age) + 0.2 (mm); accordingly, it requires an accuracy of at least 0.1 mm. However, one concern of IMT measurement is intervendor variability. In this study, we aimed to verify the intervendor variability using an IMT phantom.
Methods: An improved IMT phantom was developed, and it was possible to analyze the IMT by software for all vendors.
Results: With the vendor-specific software, the maximum difference between the devices was 0.08 mm, and the difference in quartile range was 0.06 mm. On the other hand, with the vendor-independent offline software, the maximum difference between the devices was 0.16 mm, and the quartile range of variation was 0.06 mm.
Conclusion: The intervendor variability assessed using our IMT phantom was less than 0.10 mm, and the on-board vendor-specific software was shown to reduce the difference between the devices significantly compared with the vendor-independent offline software. To further improve the vender difference, adjustment by means of vendor-specific software based on a standardized IMT phantom is warranted.
Keywords: Intervendor variability; Intima–media thickness; Quality control.
Similar articles
-
Validation of the Carotid Intima-Media Thickness Variability: Can Manual Segmentations Be Trusted as Ground Truth?Ultrasound Med Biol. 2016 Jul;42(7):1598-611. doi: 10.1016/j.ultrasmedbio.2016.02.004. Epub 2016 Apr 9. Ultrasound Med Biol. 2016. PMID: 27072077
-
Contrast-enhanced ultrasonography vs B-mode ultrasound for visualization of intima-media thickness and detection of plaques in human carotid arteries.Echocardiography. 2017 May;34(5):723-730. doi: 10.1111/echo.13513. Epub 2017 Mar 19. Echocardiography. 2017. PMID: 28317160
-
Evaluation of a new carotid intima-media thickness measurement by B-mode ultrasonography using an innovative measurement software, intimascope.Am J Hypertens. 2006 Dec;19(12):1206-12. doi: 10.1016/j.amjhyper.2006.05.010. Am J Hypertens. 2006. PMID: 17161764
-
Effects of ultrasound technology advances on measurement of carotid intima-media thickness: A review.Vasc Med. 2021 Feb;26(1):81-85. doi: 10.1177/1358863X20969826. Epub 2020 Nov 18. Vasc Med. 2021. PMID: 33203316 Review.
-
[Intima media thickness of the carotid arteries: early pointer to arteriosclerosis and therapeutic endpoint].Ultraschall Med. 2003 Jun;24(3):162-74. doi: 10.1055/s-2003-40058. Ultraschall Med. 2003. PMID: 12817310 Review. German.
Cited by
-
Coexistence of high visceral fat area and sarcopenia is associated with atherosclerotic markers in old-old patients with diabetes: A cross-sectional study.J Diabetes Investig. 2024 Oct;15(10):1510-1518. doi: 10.1111/jdi.14274. Epub 2024 Aug 6. J Diabetes Investig. 2024. PMID: 39105252 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources