. 2020 Oct;130(10):2494-2498.
doi: 10.1002/lary.28569. Epub 2020 Feb 25.

Reconfiguration of Upper Airway Stimulation Devices Utilizing Awake Endoscopy

Affiliations

Reconfiguration of Upper Airway Stimulation Devices Utilizing Awake Endoscopy

Joseph B Meleca et al. Laryngoscope. 2020 Oct.

Abstract

Objectives: Determine the rates of, reasons for, and effectiveness of device reprogramming utilizing in-office awake endoscopy (AE) in subjects who underwent upper airway stimulation (UAS) surgery.

Study design: Retrospective chart review.

Methods: Retrospective chart review at a tertiary care center on patients implanted with UAS devices from November 2015 to July 2018.

Results: Sixty patients were implanted with UAS devices. Average pre- and postsurgical total apnea-hypopnea index (AHI) was 40.6 and 4.2 with a 36.4 total AHI reduction. Treatment success based on Sher criteria was observed in 88% of patients. Postoperative titration polysomnography revealed an average functional threshold (FT) and minimum therapeutic amplitude (MTA) of 1.6 and 2.1 V, respectively. A total of 24 AEs were performed in 19 (32%) patients. The most common complaints and reasons for AE were perceived stimulus discomfort (42%), frequent awakenings (32%), and persistent fatigue or non-normalized AHI (21%). After first AE, there was a 0.87 (53%) and 0.93 (45%) V reduction in FT and MTA, respectively.

Conclusion: Patients with obstructive sleep apnea continue to succeed with UAS. Fewer than one-third of patients had postsurgical complaints requiring device reconfiguration utilizing AE to assess the pharyngeal airway and optimize device settings. Reduction in the FT and MTA after AE may improve compliance by reducing discomfort and frequent awakenings. Periodic monitoring of patients with implanted UAS devices is necessary to optimize efficacy and maintain patient compliance.

Level of evidence: 4 Laryngoscope, 130:2494-2498, 2020.

Keywords: Obstructive sleep apnea; advanced endoscopy; awake endoscopy; electrode reconfiguration; functional threshold; hypoglossal nerve; inspire device implant; minimum therapeutic amplitude; upper airway stimulation.

PubMed Disclaimer

References

BIBLIOGRAPHY

    1. He J, Kryger MH, Zorick FJ, Conway W, Roth T. Mortality and apnea index in obstructive sleep apnea. Experience in 385 male patients. Chest 1988;94:9-14.
    1. Rotenberg BW, Murariu D, Pang KP. Trends in CPAP adherence over twenty years of data collection: a flattened curve. J Otolaryngol Head Neck Surg 2016;45:43.
    1. Strollo PJ Jr, Soose RJ, Maurer JT, et al. Upper-airway stimulation for obstructive sleep apnea. N Engl J Med 2014;370:139-149.
    1. Woodson BT, Gillespie MB, Soose RJ, et al. Randomized controlled withdrawal study of upper airway stimulation on OSA: short- and long-term effect. Otolaryngol Head Neck Surg 2014;151:880-887.
    1. Boon M, Huntley C, Steffen A, et al. Upper airway stimulation for obstructive sleep apnea: results from the ADHERE registry. Otolaryngol Head Neck Surg 2018;159:379-385.

LinkOut - more resources