Effect of heparinization of catheters on pulmonary artery oximetry

Affiliations

Effect of heparinization of catheters on pulmonary artery oximetry

G M Stritter et al. J Clin Monit. 1988 Jul.

Abstract

A clinical study was performed in two phases to determine whether pulmonary artery oximeter catheters that were impregnated or bonded with heparin would affect the accuracy of measurements of in vivo mixed venous oxygen saturation (Sv-O2). In phase 1, 40 patients were catheterized with either a heparin-impregnated or a plain pulmonary artery catheter. Blood was sampled at random times to correlate in vivo with in vitro Sv-O2 measurements. In phase 2, 16 patients who were not receiving systemic heparin therapy or aspirin and who had no coagulopathies were catheterized with either a heparin-bonded or a plain pulmonary artery catheter in a blinded order. In phase 1, a total of 364 blood samples were obtained from 40 patients. Linear regression analysis of the pooled data demonstrated y = 0.98x - 0.01, r = 0.93, P less than 0.001, and n = 141 with heparin-impregnated catheters; and y = 0.87x + 8.0, r = 0.81, P less than 0.001, and n = 223 with plain catheters. The mean difference (in vivo minus in vitro) revealed a similar error (-1.3 +/- 0.4 versus -1.4 +/- 0.4, respectively, mean +/- SE). The 95% confidence limits of an individual value (+/- 8.1 versus +/- 12.3) suggested slightly greater accuracy with heparin-impregnated catheters. In phase 2, a total of 134 blood samples were obtained from 16 patients. Linear regression analysis showed nearly equal performance with heparin-bonded and plain catheters (r = 0.97 versus r = 0.98, respectively) with similar slopes (1.0 versus 1.1, respectively) but different intercepts (-0.6 versus -8.4, respectively).(ABSTRACT TRUNCATED AT 250 WORDS)

PubMed Disclaimer

Similar articles

References

    1. JAMA. 1972 Aug 14;221(7):667-73 - PubMed
    1. Anesth Analg. 1982 Aug;61(8):676-9 - PubMed
    1. AJR Am J Roentgenol. 1982 Nov;139(5):957-61 - PubMed
    1. Lancet. 1986 Feb 8;1(8476):307-10 - PubMed
    1. Can J Surg. 1982 Sep;25(5):538-43 - PubMed

LinkOut - more resources