Sutureless Closure of Arterial Cannulation Sites

Affiliations

Sutureless Closure of Arterial Cannulation Sites

Chetan Pasrija et al. Innovations (Phila). 2020 Mar/Apr.

Abstract

Objective: Percutaneous femoral cannulation for venoarterial extracorporeal membrane oxygenation (ECMO) is commonly performed but percutaneous removal of arterial cannulas has not been broadly accepted. We hypothesized that a system that allows endovascular access to ECMO circuits along with the MANTA® large-bore vascular closure device could be used to successfully close arterial ECMO cannulation sites in a large animal model.

Methods: Yorkshire swine (40 to 60 kg, n = 2) were used for this study. In the first swine, the infrarenal abdominal aorta was exposed. The aorta was cannulated once using a 15 Fr cannula and twice with a 19 Fr arterial cannula. A novel adaptor system that facilitates endovascular access to ECMO circuits was connected, and a 0.035″ Benston wire was placed through the adaptor and guided into the aorta. The cannula was removed over the wire and manual pressure was applied. The MANTA® sheath was inserted over the wire followed by the closure unit and was deployed. The process was repeated at 2 separate sites. A similar experiment was performed in a second swine, but through a median sternotomy to cannulate the ascending aorta.

Results: Good hemostasis was achieved at all cannulation sites. Angiography demonstrated unobstructed flow across all closure sites with no evidence of extravasation.

Conclusions: The data presented here support the use of the MANTA® vascular closure device for the closure of arterial cannulation sites following ECMO decannulation and demonstrates utility of a novel adaptor system for establishing endovascular access in this context.

Keywords: animal model; extracorporeal membrane oxygenation; sutureless closure.